DBS MICROPLATE, TRUNCATED V-BOTTOMED Product Number: 3033-0010
Recall
- Recall Number
- Z-0689-2013
- Event Number
- 63771
- Firm
- Perkinelmer
- FEI Number
- 3006159931
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- November 26, 2012
- Posted
- January 16, 2013
- Terminated
- March 7, 2014
- Address
- 940 Winter Street, Waltham, MA, 02451
Description
DBS MICROPLATE, TRUNCATED V-BOTTOMED Product Number: 3033-0010
Truncated V-Bottomed Microplate Wells used in assays contain defective wells.
Perkin Elmer initiated the recall on November 23, 2012. The Recall Letter will be delivered to the end customer by email facilitated through our PerkinElmer Technical Support group located in Akron Ohio. Technical Support will be asked to make sure the customers receive the Recall Letter and implement the recommended precautions: - Inspect the wells of row A and row H of the affected microplate lots for possible holes and cracks prior to use. - If a hole or crack is visually detected in a well of the microplate, the well should not be used for testing. - If a hole or crack is not visually detected in a well of the plate, it is required during use of the well, the volume of solution in the well should be visually confirmed to be consistent before and after incubation. Please contact PerkinEimer Technical Support for further information (800) 321 -9632, select prompt 2.
Nationwide Distribution.
56 kits (100 plates/kit)