FDA Recall Terminated

HTA Console System, 230 Volt, English, Product Number: 560010 (HTA system is a software-controlled hysteroscopic thermal endometrial ablation system)

Recall: Z-0688-2007 · Initiated January 24, 2007

Recall

Recall Number
Z-0688-2007
Event Number
37291
Firm
Boston Scientific Corporation One Boston
FEI Number
3001451463
Product Code
MNB
Status
Terminated
Root Cause
Other
Initiated
January 24, 2007
Posted
March 31, 2007
Terminated
December 12, 2007
Address
Scientific Place Natick, MA, 01760-1536

Description

HTA Console System, 230 Volt, English, Product Number: 560010 (HTA system is a software-controlled hysteroscopic thermal endometrial ablation system)

Reason

Defective integrated circuit board could result in loss of the system pump and patient injury (hot fluid 90 degree C into uterus) if in use during the recirculation/heating phase of treatment

Action

Boston Scientific notiifed accounts by letter dated January 24, 2007 via federal Express Mail. Accounts were requested to examine inventory, remove and return product.

Distribution

Worldwide-USA, UK, Denmark, Italy, and France

Quantity

14 units