FDA Recall Terminated

Nitinol TC Reusable Electrodes (TCN), Model Numbers: 1)TCN-10, 2) TCNK-10-R*, 3)TCN-10-3M, 4) TCNK-15, 5) TCN-15, 6) TCNK-15-C, 7) TCN-15-3M, 8) TCNK-15-R*, 9) TCN-20, 10) TCNK-20, 11) TCN-20-3M*, 12) TCNK-20-C*, 13) TCN-5, 14) TCNK-20-R*, 15) TCN-5-3M, 16) TCNK-5, 17) TCNK-10, 18) TCNK-5-C, 19) TCNK-10-C & 20) TCNK-5-R*. The TCN devices are indicated for use in radiofrequency (RF) heat lesioning of peripheral nerve tissue only.

Recall: Z-0687-2018 · Initiated July 5, 2017

Recall

Recall Number
Z-0687-2018
Event Number
78704
Firm
Cosman Medical, LLC
FEI Number
3006630150
Product Code
GXI
Status
Terminated
Root Cause
Device Design
Initiated
July 5, 2017
Terminated
September 14, 2018
Address
22 Terry Ave, Burlington, MA, 01803-2516

Description

Nitinol TC Reusable Electrodes (TCN), Model Numbers: 1)TCN-10, 2) TCNK-10-R*, 3)TCN-10-3M, 4) TCNK-15, 5) TCN-15, 6) TCNK-15-C, 7) TCN-15-3M, 8) TCNK-15-R*, 9) TCN-20, 10) TCNK-20, 11) TCN-20-3M*, 12) TCNK-20-C*, 13) TCN-5, 14) TCNK-20-R*, 15) TCN-5-3M, 16) TCNK-5, 17) TCNK-10, 18) TCNK-5-C, 19) TCNK-10-C & 20) TCNK-5-R*. The TCN devices are indicated for use in radiofrequency (RF) heat lesioning of peripheral nerve tissue only.

Reason

After multiple reprocessing cycles, the epoxy resin which holds the TCN Electrode in the hub can exhibit signs of damage. In some cases, this damage may result in the inability to fully remove blood and/or tissue residuals prior to cleaning and re-sterilizing the device.

Action

The firm, Cosman Medical, sent an "Urgent Medical Device Correction" notice and Field Correction Reply Verification Tracking Form dated July 5, 2017 to their customers. The notice described the product, problem and actions to be taken. The Customers were instructed to do the following: 1. Immediately discontinue use of devices that have already been used. Used TCN Electrodes should be disposed of in accordance with your institutions infectious material/biohazard waste control procedures. 2. Unused/un-opened TCN Electrodes should be identified as single use only (for one patient). These devices are supplied as non-sterile which require cleaning and sterilization prior to use. These TCN Electrodes may be used once the cleaning and sterilization steps are completed. After single use, the used TCN Electrode should be disposed of in accordance with your institutions infectious material/biohazard waste control procedures. 3. Complete and return the Field Correction Reply Verification Tracking Form. via Email: [email protected] or via Fax to: 781-272-6563 - Identify your email with TCN Electrodes Action in the subject line. Customers with questions may call 1-888-8COSMAN or email [email protected].

Distribution

Worldwide Distribution: US (nationwide) and country of Canada.

Quantity

11250 units (10,460 USA; 790 OUS)