FDA Recall Open, Classified

Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216

Recall: Z-0686-2025 · Initiated November 20, 2024

Recall

Recall Number
Z-0686-2025
Event Number
95847
Firm
Philips North America Llc
FEI Number
3016618143
Product Code
MHX
Status
Open, Classified
Root Cause
Under Investigation by firm
Initiated
November 20, 2024
Posted
December 17, 2024
Address
222 Jacobs St, Cambridge, MA, 02141-2289

Description

Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216

Reason

The following four issues are included: 1. Certain Channel Settings may result in an Access Point preventing a connected client device from roaming to another Access Point potentially causing the client device to appear to drop out from the network. 2. if the Access Point experiences a disconnection from the Access Point Controller exceeding 4 minutes, the Access Point may not transition through the desired state to all for clean reconnection of the client radio link. 3. Potential to experience higher than expected data dropouts, device disassociations/reassociations for their wireless clients, and frequent "Loss of AP" alerts. 4. Potential for the Smart-hopping Access Point is affected by a software issue where the Access Point will resent after 82.85 days when the frame counter reaches its limit. This causes the latest synchronization check to fail and potential for loss of patient data.

Action

Consignees were notified on about 11/20/2024 via certified mail. Consignees were instructed to identify affected units, share the communication with clinical staff, and place the Urgent Medical Device Correction with the documentation of the Philips Smart-hopping 2.0 AP 1.4 GHz and associated devices. A Philips representative will contact customers to arrange a software upgrade to E.19.02 (5.0.19.02) or greater to correct the issues. Distributors were instructed to contact customers and provide the Medical Device Correction to each.

Distribution

Nationwide distribution

Quantity

4,811 units