Seimens Mevatron M-2 6700 Medical Charged-Particle Radiation Therapy System; (Interlock #51 Jaw Field Position) Model Number 9401407, Siemens Medical Solutions, Concord, CA 94520
Recall
- Recall Number
- Z-0683-2008
- Event Number
- 46157
- Firm
- Siemens Medical Solutions USA, Inc
- FEI Number
- 2937457
- Product Code
- IYE
- Status
- Terminated
- Root Cause
- Device Design
- Initiated
- August 7, 2007
- Posted
- February 14, 2008
- Terminated
- December 1, 2010
- Address
- 4040 Nelson Ave, Concord, CA, 94520-1200
Description
Seimens Mevatron M-2 6700 Medical Charged-Particle Radiation Therapy System; (Interlock #51 Jaw Field Position) Model Number 9401407, Siemens Medical Solutions, Concord, CA 94520
Incorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in machines with serial numbers less than 3094, as they do not contain redundant sensors for jaw position.
The firm has sent consignees a Customer Safety Advisory Notice on August 7, 2007 informing them that a misplacement of the X and/or Y jaws can occur if interlock #51 Jaw Field Position is asserted and then cleared with the left-hand key rest. The firm recommends that if an interlock issue is encountered tha treatment of the patient be discontined until this issue is resolved.
Worldwide Distribution
9 units