FDA Recall
Terminated
Tarsys 1st Generation Seating System - an optional accessory for the Powered Wheelchairs.
Recall: Z-0682-03
·
Initiated March 10, 2003
Recall
- Recall Number
- Z-0682-03
- Event Number
- 25721
- Firm
- Invacare Corp
- FEI Number
- 1525712
- Product Code
- ITI
- Status
- Terminated
- Root Cause
- Other
- Initiated
- March 10, 2003
- Posted
- March 26, 2003
- Terminated
- April 13, 2012
- Address
- 1200 Taylor St, Elyria, OH, 44035
Description
Tarsys 1st Generation Seating System - an optional accessory for the Powered Wheelchairs.
Reason
A short circuit within the charger harness may cause heat damage to the units with potential for fire.
Action
On 3/10/2003, the firm mailed recall letters to all affected dealer accounts. The dealers were asked to contact all customers who have purchased powered wheelchairs with the optional Tarsys Seating System.
Distribution
The scooters were shipped to dealerships located nationwide, and to foreign distributors located in Thailand, So. Africa, Hong Kong, New Zealand, Japan, Maylasia, Korea, Isreal, Australia. 92% of affected units were shipped to domestic accounts.
Quantity
20,956 units