FDA Recall Terminated

CA 15-3 Calibrators, list 9C08-01; the pack contains six 4-mL bottles of 115D8:DF3 reactive determinants prediluted in TRIS buffer with protein stabilizers to yield the following assay values: A - 0 u/mL, B - 15 U/mL, C - 60 U/mL, D - 120 U/mL, E - 180 U/mL and F - 250 U/mL; Abbott Laboratories, Abbott Park, IL 60064

Recall: Z-0680-03 · Initiated March 4, 2003

Recall

Recall Number
Z-0680-03
Event Number
25690
Firm
Abbott Laboratories HPD/ADD
FEI Number
1415939
Product Code
MOI
Status
Terminated
Root Cause
Other
Initiated
March 4, 2003
Posted
March 26, 2003
Terminated
September 15, 2004
Address
100/200 Abbott Park Road, Abbott Park, IL, 60064

Description

CA 15-3 Calibrators, list 9C08-01; the pack contains six 4-mL bottles of 115D8:DF3 reactive determinants prediluted in TRIS buffer with protein stabilizers to yield the following assay values: A - 0 u/mL, B - 15 U/mL, C - 60 U/mL, D - 120 U/mL, E - 180 U/mL and F - 250 U/mL; Abbott Laboratories, Abbott Park, IL 60064

Reason

The CA 15-3 Calibrators may cause depressed values for the CA 15-3 High Control and for patient specimens.

Action

Recalled by letter dated 3/4/03. The accounts were informed of the depressed values for the CA 15-3 High Control and for patient specimens, and were requested to discontinue use and destroy any of the three affected lots found in inventory.

Distribution

Nationwide and Ecuador, England, Hong Kong, Chile, Japan, Canada, Singapore, Australia, Pakistan, Argentina, Thailand, Aruba, Korea and Jamaica.

Quantity

792 packs