CDI Blood Parameter Monitoring System 500 Product Usage: The CDl Blood Parameter Monitoring System 500 is intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PC02 , P02, K+, oxygen saturation, hematocrit, hemoglobin, and temperature are desired.
Recall
- Recall Number
- Z-0675-2012
- Event Number
- 60151
- Firm
- Terumo Cardiovascular Systems Corporation
- FEI Number
- 1828100
- Product Code
- DRY
- Status
- Terminated
- Root Cause
- Component change control
- Initiated
- December 6, 2011
- Posted
- January 11, 2012
- Terminated
- March 1, 2013
- Address
- 6200 Jackson Road, Ann Arbor, MI, 48103-9586
Description
CDI Blood Parameter Monitoring System 500 Product Usage: The CDl Blood Parameter Monitoring System 500 is intended for use during cardiopulmonary bypass procedures when continuous monitoring of arterial and/or venous pH, PC02 , P02, K+, oxygen saturation, hematocrit, hemoglobin, and temperature are desired.
In a recent review of in-house Service Procedures for the CDI Blood Parameter Monitoring System 500, Terumo Cardiovascular Systems (Terumo CVS) found that one of the procedures it had been performing on older systems should have been considered a recall at the time it was implemented: Terumo CVS began installing a protective shield (plastic grommet) in the system chassis during manufacture of all
Terumo sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated December 6, 2011. The letter identified the affected products, the problem, and the actions to be taken by customer/user. The letter states that Terumo CVS will update all affected CDl Monitors and no additional actions are necessary for those units previously returned that received the correction as part of the service process. For questions or concerns contact Terumo CVS Customer Service 1-800-521-2818.
Worldwide Distribution - USA (nationwide in the states of: AR, AZ, CA, FL, KS, IL, IN, LA, MA, MO, MI, MS, NJ, NV, NC, OH, PA, TN, TX, VA, WA, WI, WV including Puerto Rico and the countries of: AE, Australia, BR, Belgium, Brazil, Canada, China, Dominican Republic, EC, Germany, Guatemala, Japan, Korea, Mexico, SA, Taiwan and UY.
451 units total