Data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800) reader) for use with Diastat Anti-Beta2 Glycoprotein ELISA assay kit.

Recall: Z-0674-05 · Initiated October 26, 2004

Recall

Recall Number
Z-0674-05
Event Number
30924
Firm
Diasorin Inc.
FEI Number
2182595
Product Code
MSV
Status
Terminated
Root Cause
Other
Initiated
October 26, 2004
Posted
April 6, 2005
Terminated
July 22, 2006
Address
1951 Northwestern Ave S, Stillwater, MN, 55082-7536

Description

Data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800) reader) for use with Diastat Anti-Beta2 Glycoprotein ELISA assay kit.

Reason

Data reduction Program written for specific sutomated microtiter plate insturment (Bio-Tek ELx800 reader) is faulty in that the positive result limit is to be greater than 15 U/ml, rather than greater than or equal to 15 U/ml.

Action

Customers were sent notification letters and effectiveness check form. recommended users manually veify positve results (RSLT), using product insert specifications.

Distribution

CT, MI, MT, UT & CANADA

Quantity

7