Data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800) reader) for use with Diastat Anti-Beta2 Glycoprotein ELISA assay kit.
Recall: Z-0674-05
·
Initiated October 26, 2004
Recall
- Recall Number
- Z-0674-05
- Event Number
- 30924
- Firm
- Diasorin Inc.
- FEI Number
- 2182595
- Product Code
- MSV
- Status
- Terminated
- Root Cause
- Other
- Initiated
- October 26, 2004
- Posted
- April 6, 2005
- Terminated
- July 22, 2006
- Address
- 1951 Northwestern Ave S, Stillwater, MN, 55082-7536
Description
Data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800) reader) for use with Diastat Anti-Beta2 Glycoprotein ELISA assay kit.
Reason
Data reduction Program written for specific sutomated microtiter plate insturment (Bio-Tek ELx800 reader) is faulty in that the positive result limit is to be greater than 15 U/ml, rather than greater than or equal to 15 U/ml.
Action
Customers were sent notification letters and effectiveness check form. recommended users manually veify positve results (RSLT), using product insert specifications.
Distribution
CT, MI, MT, UT & CANADA
Quantity
7