Data reduction program written specifically for Bio-Tek ELx800 automated microtiter plate reader with use of Diastat Anti-Mitochondrial Antibody Kit
Recall: Z-0673-05
·
Initiated April 12, 2004
Recall
- Recall Number
- Z-0673-05
- Event Number
- 30922
- Firm
- Diasorin Inc.
- FEI Number
- 2182595
- Product Code
- DBM
- Status
- Terminated
- Root Cause
- Other
- Initiated
- April 12, 2004
- Posted
- April 6, 2005
- Terminated
- July 22, 2006
- Address
- 1951 Northwestern Ave S, Stillwater, MN, 55082-7536
Description
Data reduction program written specifically for Bio-Tek ELx800 automated microtiter plate reader with use of Diastat Anti-Mitochondrial Antibody Kit
Reason
The data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800 reader) was found to be faulty. This program is used to calculate the cut off values for Diastat Anti-Mitochondrial Antibody Kit. Program should include a borderline specification to indicate repeat testing is needed.
Action
A notification letter and effectiveness check form was sent to consignees. This letter recommended that users manually verify results using the specifications outlined in the package insert until a new software version is installed.
Distribution
CA, CT, KS, MA & WA
Quantity
8