FDA Recall Terminated

Data reduction program written specifically for Bio-Tek ELx800 automated microtiter plate reader with use of Diastat Anti-Mitochondrial Antibody Kit

Recall: Z-0673-05 · Initiated April 12, 2004

Recall

Recall Number
Z-0673-05
Event Number
30922
Firm
Diasorin Inc.
FEI Number
2182595
Product Code
DBM
Status
Terminated
Root Cause
Other
Initiated
April 12, 2004
Posted
April 6, 2005
Terminated
July 22, 2006
Address
1951 Northwestern Ave S, Stillwater, MN, 55082-7536

Description

Data reduction program written specifically for Bio-Tek ELx800 automated microtiter plate reader with use of Diastat Anti-Mitochondrial Antibody Kit

Reason

The data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800 reader) was found to be faulty. This program is used to calculate the cut off values for Diastat Anti-Mitochondrial Antibody Kit. Program should include a borderline specification to indicate repeat testing is needed.

Action

A notification letter and effectiveness check form was sent to consignees. This letter recommended that users manually verify results using the specifications outlined in the package insert until a new software version is installed.

Distribution

CA, CT, KS, MA & WA

Quantity

8