V6 Rear Case Assembly, Product Code 35701, a replacement part for the SIGMA Spectrum Infusion Pump -- SIGMA LLC Medina, NY 14103 --- NOTE: The speakers are a part of the rear case assembly located in the back of the infusion pump. The speaker is intended to generate audible tones during each keystroke, and during alarm conditions. However, some speakers have a variation in their internal wiring that increases the potential for the speaker to fail prematurely. Spectrum Infusion Pumps with non-functional speakers will continue to display visual alarm indications, and will otherwise continue to operate as intended. Product Code 35701 is a replacement assembly that is shipped to customers for self-repair of the Spectrum Infusion Pump. It is not a registered finished good.
Recall
- Recall Number
- Z-0672-2017
- Event Number
- 75311
- Firm
- Baxter Healthcare Corp.
- FEI Number
- 1417572
- Product Code
- FRN
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- September 28, 2016
- Terminated
- June 15, 2018
- Address
- 1 Baxter Pkwy, Deerfield, IL, 60015-4625
Description
V6 Rear Case Assembly, Product Code 35701, a replacement part for the SIGMA Spectrum Infusion Pump -- SIGMA LLC Medina, NY 14103 --- NOTE: The speakers are a part of the rear case assembly located in the back of the infusion pump. The speaker is intended to generate audible tones during each keystroke, and during alarm conditions. However, some speakers have a variation in their internal wiring that increases the potential for the speaker to fail prematurely. Spectrum Infusion Pumps with non-functional speakers will continue to display visual alarm indications, and will otherwise continue to operate as intended. Product Code 35701 is a replacement assembly that is shipped to customers for self-repair of the Spectrum Infusion Pump. It is not a registered finished good.
A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library and Rear Case Assembly in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to fail prematurely.
Baxter sent an Urgent Medical Device Correction letter dated September 28, 2016, to all affected customers. On 10/27/2016, Baxter Healthcare sent a f/u UMDC Letter (dated 10/27/2016) to customers. The f/u letter contains the device correction plan which states that Baxter will be replacing at no charge the rear case assembly on all affected Sigma Spectrum Infusion Pumps, as well as affected stand-alone rear case assemblies shipped to customers as spare parts. For questions, please contact your Baxter sales representative, or Baxter Healthcare Medina at 800-356-3454 (choose option 1) Monday through Friday, between 7:00 am and 7:00 pm ET.
Nationwide Distribution, Puerto Rico and Canada
439 units