FDA Recall Terminated

Bivona Uncuffed Neonatal Tracheostomy Tube Sizes 2.5 - 4.0 (With a Straight Neck Flange) Product Usage: This tube is intended to provide direct airway access for a tracheostomized patient for up to 29 days. It may be reprocessed up to 5 times for single patient reuse.

Recall: Z-0671-2012 · Initiated November 28, 2011

Recall

Recall Number
Z-0671-2012
Event Number
60449
Firm
Smiths Medical ASD, Inc.
FEI Number
1824231
Product Code
BTO
Status
Terminated
Root Cause
Device Design
Initiated
November 28, 2011
Posted
January 19, 2012
Terminated
October 17, 2014
Address
5700 W 23rd Ave, Gary, IN, 46406-2617

Description

Bivona Uncuffed Neonatal Tracheostomy Tube Sizes 2.5 - 4.0 (With a Straight Neck Flange) Product Usage: This tube is intended to provide direct airway access for a tracheostomized patient for up to 29 days. It may be reprocessed up to 5 times for single patient reuse.

Reason

Difficulty disconnecting accessories from the Bivona Uncuffed Neonatal, Pediatric and Flextend Tracheostomy Tubes TR3 connector, which may result in a delay in initiating therapy or patient discomfort. If excessive force is used to detach the accessory, this may result in decannulation of the tube which could result in serious injury or death; especially if no replacement tracheostomy tube is imm

Action

Smith Medical sent an Urgent Field Safety Notice dated November 28, 2011, via U.S. Mail to all affected consignees. The notice identified the product, description of problem, and actions to be taken. This notice included a Customer Information Bulletin and Urgent Field Safety Notice Confirmation Form. Customers were instructed to 1) identify all affected unused product in inventory and segregate it to a quarantine location; 2) complete and return the attached Confirmation Form by fax to 219-989-7259 or email to [email protected]; and 3) distributors were instructed to stop distributing and provide Smith Medical with a list of their customers, including contact details, so that Smith Medical can notify the customers of this notice. For questions contact Smiths Medical Customer Service at 800-258-5361, Option 1.

Distribution

Worldwide Distribution - USA Nationwide: AZ, CA, CO, CT,DE, FL, GA, HI, IL, IN, KS, KY, LA, SC, TX, OR, NJ, MO, UT, MI, NY, NM, MA, MN, MS, NC, ND, NE, NH, NV, OH, OK, OR, RI, SD, TN, VA, VT, WA, WI, WV including Washinton DC, Australia, Austria, Barbados, Belgium, Brazil, Canada, Chile, Cyprus, Czech Republic, Denmark, Egypt, Finland, France, Germany, Gibraltar, Great Britain (UK), Greece, Hong Kong, Ireland, Israel, Italy, Kuwait, Libya, Malaysia, Netherlands, New Zeland, Norway, Poland, Portugal, Qatar, Russian Federation, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, United Arab Emirates, United States, Vatican City State

Quantity

US distribution 153,033 devices ; Foreign 21,490 devices