FDA Recall Terminated

sigma spectrum VOLUMETRIC INFUSION PUMP with Master Drug Library, V6 Spectrum Infusion Pump, Product Code 35700BAX and 35700BAXR, Unique Device Identifier GTIN 00085412091570, Rx ONLY -- Sigma, LLC Medina, NY 14103 --- NOTE: The speakers are a part of the rear case assembly located in the back of the infusion pump. The speaker is intended to generate audible tones during each keystroke, and during alarm conditions. However, some speakers have a variation in their internal wiring that increases the potential for the speaker to fail prematurely. Spectrum Infusion Pumps with non-functional speakers will continue to display visual alarm indications, and will otherwise continue to operate as intended.

Recall: Z-0670-2017 · Initiated September 28, 2016

Recall

Recall Number
Z-0670-2017
Event Number
75311
Firm
Baxter Healthcare Corp.
FEI Number
1417572
Product Code
FRN
Status
Terminated
Root Cause
Process control
Initiated
September 28, 2016
Terminated
June 15, 2018
Address
1 Baxter Pkwy, Deerfield, IL, 60015-4625

Description

sigma spectrum VOLUMETRIC INFUSION PUMP with Master Drug Library, V6 Spectrum Infusion Pump, Product Code 35700BAX and 35700BAXR, Unique Device Identifier GTIN 00085412091570, Rx ONLY -- Sigma, LLC Medina, NY 14103 --- NOTE: The speakers are a part of the rear case assembly located in the back of the infusion pump. The speaker is intended to generate audible tones during each keystroke, and during alarm conditions. However, some speakers have a variation in their internal wiring that increases the potential for the speaker to fail prematurely. Spectrum Infusion Pumps with non-functional speakers will continue to display visual alarm indications, and will otherwise continue to operate as intended.

Reason

A correction was issued for the SIGMA Spectrum Infusion Pump with Master Drug Library in order to replace speakers installed in pumps or included in spare part rear case assemblies that have an increased potential to fail prematurely.

Action

Baxter sent an Urgent Medical Device Correction letter dated September 28, 2016, to all affected customers. On 10/27/2016, Baxter Healthcare sent a f/u UMDC Letter (dated 10/27/2016) to customers. The f/u letter contains the device correction plan which states that Baxter will be replacing at no charge the rear case assembly on all affected Sigma Spectrum Infusion Pumps, as well as affected stand-alone rear case assemblies shipped to customers as spare parts. For questions, please contact your Baxter sales representative, or Baxter Healthcare Medina at 800-356-3454 (choose option 1) Monday through Friday, between 7:00 am and 7:00 pm ET.

Distribution

Nationwide Distribution, Puerto Rico and Canada

Quantity

5,976 units (US: 5972, Canada: 4)