FDA Recall Terminated

Disposable components of the superDimension Bronchus System: Locatable Guide Kit (AKI00017-01) and Extended Working Channel (AKI00041-01).

Recall: Z-0667-2007 · Initiated November 29, 2006

Recall

Recall Number
Z-0667-2007
Event Number
36947
Firm
SuperDimension, Inc.
FEI Number
3004962788
Product Code
JAK
Status
Terminated
Root Cause
Other
Initiated
November 29, 2006
Posted
March 29, 2007
Terminated
November 20, 2008
Address
161 Cheshire Lane North, Suite 100, Minneapolis, MN, 55441

Description

Disposable components of the superDimension Bronchus System: Locatable Guide Kit (AKI00017-01) and Extended Working Channel (AKI00041-01).

Reason

The action is being taken in response to certain incidents that have been reported to the company, one of which concerned an adverse event in which a patient incurred a pneumothorax during a procedure in which a superDimension Bronchus System was used. The biopsy results for the patient were taken from a different area than the confirmed navigation site that was indicated by the system.

Action

An Urgent Medical Device Field Action letter, dated 11/30/2006, was sent to medical facilities who have the superDimension Bronchus System installed. The letter infoms the facility to immediately stop using the Bronchus System, and move all inventory of the disposable system components to a controlled quarantine area. A superDimension representative will contact the consignees to schedule a time to retrieve the quarantined inventory, at which time a verification form will be completed.

Distribution

Worldwide, including USA, Austria, Belgium, Canada, France, Germany, Hungary, Israel, Italy, Netherlands, South Africa, Spain, Switzerland, Turkey, and United Kingdom.

Quantity

Up to 2,000 each Locatable Guide Kit and Extended Working Channel