FDA Recall Open, Classified

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Recall: Z-0664-2025 · Initiated May 31, 2023

Recall

Recall Number
Z-0664-2025
Event Number
95603
Firm
Micro-X Ltd. Unit 14 6 Mab Eastern Promenade Tonsley Australia
FEI Number
3012855798
Product Code
IZL
Status
Open, Classified
Root Cause
Software design
Initiated
May 31, 2023
Posted
December 12, 2024

Description

MICRO-X Rover Mobile X-ray System, # MXU-RV35

Reason

Shots were terminated by the mAs integrator; however, the 4 ms exposure time did not allow enough tolerance to achieve diagnostic exposures due to variations in hardware and how different timers are started.

Action

The firm will notify consignees using a service bulletin that was already posted. Additionally, the firm will implement a firmware update to extend the timer (4 ms to 10 ms minimum) to ensure even short exposures achieve the necessary image quality for diagnostic use.

Distribution

Worldwide and US Nationwide Distribution: NJ, PR, CA

Quantity

5