FDA Recall Terminated

ASPIRE Cristalle (for FDR-3000AWS) The Fujifilm Digital Mammography System, ASPIRE Cristalle (FDR MS-3500) generates full field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.

Recall: Z-0661-2018 · Initiated August 4, 2017

Recall

Recall Number
Z-0661-2018
Event Number
79078
Firm
Fujifilm Medical Systems U.S.A., Inc.
FEI Number
1000513161
Product Code
MUE
Status
Terminated
Root Cause
Software design
Initiated
August 4, 2017
Terminated
April 1, 2019
Address
419 West Ave, Stamford, CT, 06902-6343

Description

ASPIRE Cristalle (for FDR-3000AWS) The Fujifilm Digital Mammography System, ASPIRE Cristalle (FDR MS-3500) generates full field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.

Reason

FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR1000AWS, FDR-2000AWS, FDR-3000AWS, and CRIR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may overwrite the image already stored on PACS.

Action

The firm, FUJIFILM Medical Systems U.S.A., Inc. (FMSU), sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated 7/28/2017 to its customers. The letter described the products, problem and actions to be taken. The customers were instructed to please read and follow the instructions in the "ACTIONS TO BE TAKEN BY CUSTOMER/USER" section. If you have an error, contact your local FUJIFILM office; and complete and return the Customer Feedback Form. FUJIFILM service personnel will contact all of the medical facilities where the applicable products have been installed to arrange for this correction and visit to take corrective measures.. If you have any questions about this matter, call Recall contact (203) 276-3445 or email: [email protected].

Distribution

Worldwide Distribution.

Quantity

71 units (68 units Domestic. 3 units Foreign) in total