Rainbow DC-I-DC Pulse CO-Oximeter Sensor used with the Masimo SET Rad 57 Pulse CoOximeter
Recall
- Recall Number
- Z-0661-06
- Event Number
- 34569
- Firm
- Masimo Corporation
- FEI Number
- 3001452206
- Product Code
- DQA
- Status
- Terminated
- Root Cause
- Other
- Initiated
- February 2, 2006
- Posted
- March 18, 2006
- Terminated
- June 17, 2008
- Address
- 40 Parker, Irvine, CA, 92618-1604
Description
Rainbow DC-I-DC Pulse CO-Oximeter Sensor used with the Masimo SET Rad 57 Pulse CoOximeter
Internal testing revealed that some of the sensors might not meet the accuracy specifications of : 2% for SpO2 and of: 3% for SpCO, due to the tolerance of one of the emitter LED's used in the calculation ofSpO2 and SpCO. The SpO2 reading may have a negative bias of up to 1 % and the SpCO reading may have a negative bias of up to 4%.
Masimo notified all customers by letter sent via Federal Express who were shipped Rainbow DC- I-DC sensors to return them to Masimo for repair or replacement. A Tracking/Verification form will be mailed with the letter to be faxed back to Masimo. Masimo will repair or replace the returned units and return them back to the customer.
Nationwide and Belgium, Canada, France, Germany, Italy, South Korea, new Zealand, Slovenia, Spain, Sweden, Switzerland, and United Arab Emirates
321