Optisure Dual Coil Defibrillation Leads, Model No. LDA220, LDA220Q, LDA230Q, LDP220Q. St. Jude Medical. Designed for long term attachment to an Implantable Cardioverter Defibrillator (ICD).
Recall
- Recall Number
- Z-0660-2016
- Event Number
- 72569
- Firm
- St Jude Medical Cardiac Rhythm Management Division
- FEI Number
- 2017865
- Product Code
- LWS
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- November 3, 2015
- Posted
- January 21, 2016
- Terminated
- May 10, 2019
- Address
- 15900 Valley View Ct, Sylmar, CA, 91342-3577
Description
Optisure Dual Coil Defibrillation Leads, Model No. LDA220, LDA220Q, LDA230Q, LDP220Q. St. Jude Medical. Designed for long term attachment to an Implantable Cardioverter Defibrillator (ICD).
Leads may have been compromised during the manufacturing process, with cuts in the insulation layer of the SVC shock coil.
An Important Medical Device Advisory letter was sent to doctors on 11/3/15 to inform them of the recall by St. Jude Medical. The letter informs the doctors that St. Jude Medical identified that during the manufacturing process of a limited number of Optisure leads, a trim technique to remove excess medical adhesive around the SVC shock coil may have introduced damage to the leads insulation. The letter provides the doctors with the recommendations for patients with a potentially impacted Optisure lead connected to a device with Dynamic TX technology and for patients implanted with a potentially impacted Optisure lead connected to a device without DynamicTx technology. Doctors with questions about the advisory are instructed to contact their local Sales Representative or St. Jude Medical Technical Services at (800) 722-3774. A revised dear doctor letter was sent via FedEx on 1/21/16 to all affected physicians. A press release will be sent/posted on 1/22/16.
Worldwide Distribution US (nationwide), Belgium, Switzerland, Germany, France, United Kingdom, India, Italy, Japan, Luxembourg, The Netherlands, Saudi Arabia, and Sweden.
447 units total (281 units in US)