FDA Recall
Terminated
Spacelabls Medical, Ultraview Multigas Analyzer, Model 91518,Spacelabs Healthcare, Issaquah, WA, 98027
Recall: Z-0660-2008
·
Initiated October 8, 2007
Recall
- Recall Number
- Z-0660-2008
- Event Number
- 44965
- Firm
- Spacelabs Healthcare, Incorporated
- FEI Number
- 3010157426
- Product Code
- CBR
- Status
- Terminated
- Root Cause
- Software design
- Initiated
- October 8, 2007
- Posted
- January 31, 2008
- Terminated
- April 12, 2011
- Address
- 5150 220th Ave Se, Issaquah, WA, 98029-6834
Description
Spacelabls Medical, Ultraview Multigas Analyzer, Model 91518,Spacelabs Healthcare, Issaquah, WA, 98027
Reason
Inaccurate partial pressure readings; readings are not compensated for ambient pressure in mmHg or kPa.
Action
The firm sent out their Urgent Medical Device correction/recall letters on 10/5/2007, and an updated versions on 11/30/07 and 12/7/2007, forwarding an attached set of work around directions for use and associated tables for calculations. This short term fix will be updated to a permanent fix when a new software revision is released and installed at no cost to each user as noted in the recall notification. Telephone contact was provided at 1-800-522-7025.
Distribution
Worldwide
Quantity
722 devices