FDA Recall Terminated

Baxter SPS 550 Single Patient System Hemodialysis Machine; Baxter Healthcare Corporation, Renal Division, 1425 Lake Cook Road, Deerfield, IL 60015

Recall: Z-0660-03 · Initiated February 20, 2003

Recall

Recall Number
Z-0660-03
Event Number
25607
Firm
Baxter Healthcare Corp. Rt
FEI Number
1416980
Product Code
FKP
Status
Terminated
Root Cause
Other
Initiated
February 20, 2003
Posted
March 20, 2003
Terminated
March 27, 2006
Address
120 & Wilson Rd, Round Lake, IL, 60073

Description

Baxter SPS 550 Single Patient System Hemodialysis Machine; Baxter Healthcare Corporation, Renal Division, 1425 Lake Cook Road, Deerfield, IL 60015

Reason

Excess or insufficient ultrafiltration may occur during manual control of ultrafiltration via adjustment of transmembrane pressure (TMP)

Action

Baxter sent Safety Alert letters dated 2/20/03 to the end users, informing them of the possibility of either excess or insufficient ultrafiltration when the operator manually controls ultrafiltration via adjustment of transmembrane pressure. The accounts were informed that turning off the UF controller during fault codes FL-04, FL-07 and FL-08 may affect the ability to safely perform manual ultrafiltration, and that treatment should be discontinued if those three fault codes are displayed. Changes to the instructions for use are in process and will be sent to the end users when completed. Additionally, the Baxter 1550 has a side panel that contains a 16-digit display. Currently, during all ultrafiltration system fault modes the display indicates 'Manual UF Control'. To avoid operator confusion, new software is being developed that will provide a display of 'Discontinue TX' when a FL-04, FL-07 or FL-08 fault code occurs. The SPS 550 does not have the side panel display and does not require the software upgrade.

Distribution

Nationwide and internationally to Algeria, AMerican Samoa, Argentina, Aruba, Australia, Barbados, Belgium, Bolivia, Brazil, Brunei, Burma, Canada, Chile, China, Colombia, Croatia, Cyprus, Dominican Republic, Ecuador, Egypt, El Salvador, Yugoslavia, Germany, Ghana, Guam, Guatemala, Honduras, Hong Kong, Hungary, INdia, Indonesia, Israel, Italy, Jamaica, Jordan, Lebanon, Malaysia, Mexico, Nicaragua, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Romania, Russia, South Korea, Singapore, Slovakia, Slovenia, South Africa, Spain, St. Maarten, Taiwan, Thailand, Tobago, Trinidad, Turkey, United Kingdom, Uruguay, Venezuela and Vietnam.