FDA Recall Terminated

AutoPulse Resuscitation System Model 100, Mfg by:Zoll Circulation Sunnyvale, CA. The AutoPulse Model 100 Automatic Mechanical Chest compressor is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.

Recall: Z-0658-2013 · Initiated August 8, 2012

Recall

Recall Number
Z-0658-2013
Event Number
62950
Firm
Zoll Circulation, Inc.
FEI Number
3003793491
Product Code
DRM
Status
Terminated
Root Cause
Employee error
Initiated
August 8, 2012
Posted
January 9, 2013
Terminated
May 12, 2014
Address
650 Almanor Ave, Sunnyvale, CA, 94085-3513

Description

AutoPulse Resuscitation System Model 100, Mfg by:Zoll Circulation Sunnyvale, CA. The AutoPulse Model 100 Automatic Mechanical Chest compressor is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.

Reason

Complaints associated with an unexpected stop in AutoPulse compressions.

Action

Zoll sent Customer Notification letters dated August 8, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Zoll Technical Support Line at 1 (800) 348-9011 for further assistance.

Distribution

Worldwide distribution.

Quantity

3881 in US, 1846 ROW.