FDA Recall
Terminated
AutoPulse Resuscitation System Model 100, Mfg by:Zoll Circulation Sunnyvale, CA. The AutoPulse Model 100 Automatic Mechanical Chest compressor is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.
Recall: Z-0658-2013
·
Initiated August 8, 2012
Recall
- Recall Number
- Z-0658-2013
- Event Number
- 62950
- Firm
- Zoll Circulation, Inc.
- FEI Number
- 3003793491
- Product Code
- DRM
- Status
- Terminated
- Root Cause
- Employee error
- Initiated
- August 8, 2012
- Posted
- January 9, 2013
- Terminated
- May 12, 2014
- Address
- 650 Almanor Ave, Sunnyvale, CA, 94085-3513
Description
AutoPulse Resuscitation System Model 100, Mfg by:Zoll Circulation Sunnyvale, CA. The AutoPulse Model 100 Automatic Mechanical Chest compressor is intended to be used as an adjunct to manual CPR, on adult patients only, in cases of clinical death as defined by lack of spontaneous breathing and pulse.
Reason
Complaints associated with an unexpected stop in AutoPulse compressions.
Action
Zoll sent Customer Notification letters dated August 8, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Zoll Technical Support Line at 1 (800) 348-9011 for further assistance.
Distribution
Worldwide distribution.
Quantity
3881 in US, 1846 ROW.