FDA Recall Terminated

Reliance LITe Nerve Hook Bayoneted, 90, Blunt Tip, REF # 48066103 The blunt probe can be used as a conventional means of achieving nerve retraction. The bend at the proximal tip of the ring the discectomy, as well as during the insertion of an IBD during a conventional TLIF procedure.

Recall: Z-0657-2019 · Initiated November 26, 2018

Recall

Recall Number
Z-0657-2019
Event Number
81645
Firm
Howmedica Osteonics Corp.
FEI Number
3004024955
Product Code
LXH
Status
Terminated
Root Cause
Under Investigation by firm
Initiated
November 26, 2018
Terminated
May 26, 2020
Address
2 Pearl Ct, Allendale, NJ, 07401-1611

Description

Reliance LITe Nerve Hook Bayoneted, 90, Blunt Tip, REF # 48066103 The blunt probe can be used as a conventional means of achieving nerve retraction. The bend at the proximal tip of the ring the discectomy, as well as during the insertion of an IBD during a conventional TLIF procedure.

Reason

Instead of having a spherical diameter, the probe was observed to have 2 radii and a flat surface. Additionally, upon further investigation, it was observed that some parts exhibited an edge on the intersection between the radius and the diameter of the probe.

Action

The firm, Stryker, sent an "Urgent Medical Device Recall" letter dated 11/14/2018 to its customers on 11/26/2018. The letter described product, problem and actions to be taken. The customers were instructed to: Examine your inventory; Immediately quarantine the affected product; and complete customer response form to Stryker Spine at email: [email protected]. For questions contact Manager-Regulatory Compliance & Quality System at 866-45-SPINE (77463)..

Distribution

Worldwide distribution: US Distribution to the state of: PA and countries of:: Australia, Canada, France, Italy, Netherlands, Poland, and Spain.

Quantity

39