FDA Recall Open, Classified

Abiomed KIT, 14 Fr Introducer, 13 & 25 cm, ABMD Part of Pump Set (0048-0024, 0048-0032)-Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 0052-3056

Recall: Z-0654-2024 · Initiated November 7, 2023

Recall

Recall Number
Z-0654-2024
Event Number
93501
Firm
Abiomed, Inc.
FEI Number
1220648
Product Code
DYB
Status
Open, Classified
Root Cause
Process control
Initiated
November 7, 2023
Posted
January 5, 2024
Address
24 Cherry Hill Dr, Danvers, MA, 01923-2575

Description

Abiomed KIT, 14 Fr Introducer, 13 & 25 cm, ABMD Part of Pump Set (0048-0024, 0048-0032)-Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 0052-3056

Reason

Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis

Action

Abiomed issued URGENT MEDICAL DEVICE RECALL (REMOVAL) Letter Products Impacted: Introducer Kit for Impella (sold within Impella Pump sets and Individually Packaged) Axillary Introducer Kits on 11/7/23. Letter states reason for recall, health risk and action to take: IF AN IMPELLA PUMP IS CLINICALLY NEEDED, ACTIONS TO BE TAKEN BY THE USER: 1. Prior to usage, carefully examine the outer pouch of the Introducer Kit for holes following the provided instructions in Attachment 2. 2. For non-trayed introducers (Part Numbers 0052-3006): IF THERE IS ANY SUSPICION OF THIS TYPE OF PACKAGING DAMAGE, REFRAIN FROM USING THE PRODUCT. 3. For trayed introducers with an inner tray package (Part Numbers 0052-3014, 0052-3015, 0052-3021, 0052-3052, 0052-3046, 0052-3025, 0052-3056, 2000342, 0052-3053): Consider that while the outside surface of the tray is non-sterile, the introducers inside the tray are considered sterile. After removing the inner tray from the outer pouch, avoid introducing it to the sterile field. Instead, employ sterile technique to open the inner tray, preserving its interior sterility. a. Transfer the contents from the inner tray to the sterile field following your institution s sterile field management protocol, ensuring no contact with the outer surface to prevent contamination. b. The inner tray must remain separate from the sterile field and should be discarded once all components of the introducer kit have been extracted. 4. If used, document the use on the Business Request Form at the end of this letter. Please contact Abiomed s Clinical Support Center, 1-800-422-8666, if you need any support troubleshooting or if you have any further questions.

Distribution

Nationwide Foreign: Country AT AU BE BR CA CA CH CZ DE DK ES FR FR HK IN IT MX MY NL NO PA SG TW UK

Quantity

2 units within kit