FDA Recall Terminated

Vantage Panoramic X-Ray System used to perform dental and orthodontic radiological imaging The Vantage Panoramic X-Ray System is a class II extra oral X-Ray system used to provide dental radiographic examination and diagnosis of diseases of the teeth, jaw and oral structures. When the system is equipped with the cephalometric option, the system will also provide cephalometric radiographic examinations for use in orthodontic treatment planning and evalu

Recall: Z-0651-2016 · Initiated September 4, 2014

Recall

Recall Number
Z-0651-2016
Event Number
70765
Firm
Midmark Corp dba Progeny Inc
FEI Number
3000210017
Product Code
MUH
Status
Terminated
Root Cause
Device Design
Initiated
September 4, 2014
Terminated
December 29, 2017
Address
675 Heathrow Dr, Lincolnshire, IL, 60069-4206

Description

Vantage Panoramic X-Ray System used to perform dental and orthodontic radiological imaging The Vantage Panoramic X-Ray System is a class II extra oral X-Ray system used to provide dental radiographic examination and diagnosis of diseases of the teeth, jaw and oral structures. When the system is equipped with the cephalometric option, the system will also provide cephalometric radiographic examinations for use in orthodontic treatment planning and evalu

Reason

Failure of the primary collimator may result in unintentional movement of the device. The unintentional movement will not result in the generation of an error message.

Action

MidMark sent an Urgent Medical Device Field Correction letter dated September 4, 2014, to all affected customers. The letter included instructions for distributors to: 1) encourage end users to contact Midmark to arrange the field correction after the final solution becomes available; 2) if a partial image is reported to have occurred with a machine, the machine should be removed from use and the end user should initiate a service request per the normal procedure; 3) quarantine any affected machines in the distributor's inventory and contact Midmark Technical Support within 10 business days to provide the Serial numbers to on-site servicing of the affected machines can be scheduled. Distributors with questions about this market action can contact Midmark Technical Support by calling 888-294-3800 x 3 or by e-mail at [email protected].

Distribution

Worldwide Distribution - US including AL, AR, AZ, CO, CT, DE, FL, GA, IA, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, NE, NJ, NM, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI; and Internationally to Australia, Belgium, Canada, Czech Republic, France, Hungary, Lithuania, New Zealand, Poland, Russia, Saudi Arabia, South Africa, Switzerland, Turkey, United Arab Emirates, United Kingdom, and Vietnam.

Quantity

832 systems