Nova StatStrip/StatSensor Hospital Blood Glucose Meter; Nova StatStrip Lactate Hospital Meter System; Nova StatSensor Creatinine Hospital Meter, Rechargeable Lithium Batteries Single: 46626; 4 Pack: 46827; 5 Pack: 50436. For Point-Of-Care usage in the quantitative determination of Glucose in whole blood.

Recall: Z-0650-2014 · Initiated October 10, 2013

Recall

Recall Number
Z-0650-2014
Event Number
66398
Firm
Nova Biomedical Corporation
FEI Number
1219029
Product Code
CGA
Status
Terminated
Root Cause
Component design/selection
Initiated
October 10, 2013
Posted
January 9, 2014
Terminated
March 24, 2014
Address
200 Prospect St, Waltham, MA, 02453

Description

Nova StatStrip/StatSensor Hospital Blood Glucose Meter; Nova StatStrip Lactate Hospital Meter System; Nova StatSensor Creatinine Hospital Meter, Rechargeable Lithium Batteries Single: 46626; 4 Pack: 46827; 5 Pack: 50436. For Point-Of-Care usage in the quantitative determination of Glucose in whole blood.

Reason

Revised Battery Instructions for Removal and Expiration Date due to battery swelling and cell leakage

Action

On November 18, 2013, an URGENT Field Correction Letter was sent to all StatStrip Glucose Hospital meter, StatStrip Lactate Hospital Meter and StatSensor Creatinine Hospital Meter customers via email to the Point of Care Coordinator to alert users that lithium batteries that are utilized beyond their intended useful life may swell and/or crack posing a potential safety hazard. In some cases, the swelling may potentially cause the battery door on the meter to dislodge and render the meter inoperable.

Distribution

Nationwide Distribution.

Quantity

est. 200,000