Stryker Medical Stair-PRO Evacuation Chair civilian use stair chair, patient transport device, Model 6253.
Recall: Z-0650-2009
·
Initiated November 18, 2008
Recall
- Recall Number
- Z-0650-2009
- Event Number
- 50178
- Firm
- Stryker Medical Div. of Stryker Corporation
- FEI Number
- 1831750
- Product Code
- FPP
- Status
- Terminated
- Root Cause
- Process control
- Initiated
- November 18, 2008
- Posted
- February 4, 2009
- Terminated
- October 27, 2010
- Address
- 3800 E. Centre Ave., Portage, MI, 49002
Description
Stryker Medical Stair-PRO Evacuation Chair civilian use stair chair, patient transport device, Model 6253.
Reason
Handle could come off. The bolts may back out over time, resulting in a risk of dropping the patient and/or injury to transport personnel, a delay in medical treatment, or a delay in evacuation.
Action
Consignees were notified by Urgent Medical Device Correction letter dated 11/18/08 and given the serial numbers of their units, an upgrade kit and instructions for performing replacement of the bolts. Consignees were informed Stryker will perform the upgrade if requested.
Distribution
Worldwide Distribution: USA, India, Netherlands and United Kingdom.
Quantity
102