FDA Recall Terminated

iChem 10 SG Urine Chemistry Strips, PN 800-7005. Intended for the in vitro measurement of glucose, protein, bilirubin, urobilinogen, pH, specific gravity, blood, ketones, nitrite, leukocyte esterase, ascorbic acid, and color.

Recall: Z-0649-2016 · Initiated December 8, 2014

Recall

Recall Number
Z-0649-2016
Event Number
72925
Firm
Beckman Coulter Inc.
FEI Number
2050012
Product Code
JIL
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
December 8, 2014
Terminated
January 13, 2016
Address
250 S Kraemer Blvd, Brea, CA, 92821-6232

Description

iChem 10 SG Urine Chemistry Strips, PN 800-7005. Intended for the in vitro measurement of glucose, protein, bilirubin, urobilinogen, pH, specific gravity, blood, ketones, nitrite, leukocyte esterase, ascorbic acid, and color.

Reason

Iris International is recalling the iChem 10 SG Urine Chemistry Strip because of false positive leukocyte result for samples that are truly negative.

Action

An "Important Product Notice," dated 12/8/14 was sent to customers to inform them of the recall by Iris International. The recall letter informs customers that The identified lots of iChem 10 SG Urine Chemistry Strips may have leukocyte pads that are more sensitive at the first reporting range of 1+, corresponding to 25 white blood cells (WBCs) /microliter. The letter provides the impact of the recall and the actions to taken by the customers. Customers are instructed to discard any remaining strips from the affected lots and use another lot number. Customers can either contact their local Customer Service Support or complete and return the Replacement Order form, if replacement product is needed. The enclosed response form should be completed and returned within 10 days. Customers with questions are instructed to refer to http://www.beckmancoulter.com; (800) 854-3633 in US and Canada. Customers outside the US and Canada are instructed to contact their local support representative.

Distribution

Worldwide Distribution -- US, Brazil, Canada, Chile, Ecuador, France, Germany, Greece, Italy, Kuwait, Luxembourg, Malaysia, Philippines, Portugal, Puerto Rico, Singapore, Spain, Switzerland, Turkey, and United Arab Emirates.

Quantity

16,206 vials total (7,827 vials in US)