Immunoradiometric assay. Device is an immunoradiometric assay designed for the quantitative measurement of thyroid stimulating hormone (thyrotropin, TSH) in serum. It is intended strictly for in vitro diagnostic. Used as an aid in the assessment of thyroid status.
Recall
- Recall Number
- Z-0648-2013
- Event Number
- 63852
- Firm
- Siemens Healthcare Diagnostics
- FEI Number
- 2432235
- Product Code
- JLW
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- November 28, 2012
- Posted
- January 7, 2013
- Terminated
- October 28, 2013
- Address
- 511 Benedict Ave, Tarrytown, NY, 10591-5005
Description
Immunoradiometric assay. Device is an immunoradiometric assay designed for the quantitative measurement of thyroid stimulating hormone (thyrotropin, TSH) in serum. It is intended strictly for in vitro diagnostic. Used as an aid in the assessment of thyroid status.
The recall was initiated because Siemens Healthcare Diagnostics confirmed an under-recovery of dose values with some vials of Coat-A-Consta (CAC) IRMA TSH calibrators TSI4 (level B), TSI5 (level C), TSI6 (level D), TSI8 (level F), and TSI9 (level G) lot 035. The affected calibrator vials can be identified by a darker yellow color than normal. The TSI3 (level A), TSI7 (level E), and TSIX (level H)
Siemens sent an Urgent Medical Device Recall letter dated November 2012 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product, complete and return the Urgent Medical Device Recall Effectiveness Check form included with the letter by fax to 302-631-7597. Customers were also instructed to determine their replacements needed, and to forward the recall letter to whomever they may have distributed the affected product. For questions customers were instructed to contact their Siemens Technical Solutions Ceter or their local technical support representative. For questions regarding this recalla call 914-524-2955.
Worldwide Distribution - USA including CA, GA, IL, IN, MI, MN, NY, and Puerto Rico. Internationally to Agentina, Bolivia, Brazil, Canada, Chile, Guatemala, South Korea, Philippines, Paraguay, Taiwan, and Uraguay,
Lot 742= 8 units all foreign consignees) Lot 743=7 units (all foreign consignees)