FDA Recall Terminated

Immunoradiometric assay.

Recall: Z-0647-2013 · Initiated November 28, 2012

Recall

Recall Number
Z-0647-2013
Event Number
63852
Firm
Siemens Healthcare Diagnostics
FEI Number
2432235
Product Code
JLW
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
November 28, 2012
Posted
January 7, 2013
Terminated
October 28, 2013
Address
511 Benedict Ave, Tarrytown, NY, 10591-5005

Description

Immunoradiometric assay.

Reason

The recall was initiated because Siemens Healthcare Diagnostics confirmed an under-recovery of dose values with some vials of Coat-A-Consta (CAC) IRMA TSH calibrators TSI4 (level B), TSI5 (level C), TSI6 (level D), TSI8 (level F), and TSI9 (level G) lot 035. The affected calibrator vials can be identified by a darker yellow color than normal. The TSI3 (level A), TSI7 (level E), and TSIX (level H)

Action

Siemens sent an Urgent Medical Device Recall letter dated November 2012 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product, complete and return the Urgent Medical Device Recall Effectiveness Check form included with the letter by fax to 302-631-7597. Customers were also instructed to determine their replacements needed, and to forward the recall letter to whomever they may have distributed the affected product. For questions customers were instructed to contact their Siemens Technical Solutions Ceter or their local technical support representative. For questions regarding this recalla call 914-524-2955.

Distribution

Worldwide Distribution - USA including CA, GA, IL, IN, MI, MN, NY, and Puerto Rico. Internationally to Agentina, Bolivia, Brazil, Canada, Chile, Guatemala, South Korea, Philippines, Paraguay, Taiwan, and Uraguay,

Quantity

Lot 831=19 units (all foreign consignees) and Lot 832=2 units (all foreign consignees)