FDA Recall Terminated

Heartsine Samaritan AED Defibrillators - Models: SAM 001, SAM 002, SAM 003. Serial numbers within range of 1270 - 2324 (not all are affected).

Recall: Z-0641-05 · Initiated February 14, 2005

Recall

Recall Number
Z-0641-05
Event Number
31212
Firm
Heartsine Technologies Inc
FEI Number
3004018643
Product Code
MKJ
Status
Terminated
Root Cause
Nonconforming Material/Component
Initiated
February 14, 2005
Posted
March 23, 2005
Terminated
June 8, 2011
Address
940 Calle Amanecer, Ste E, San Clemente, CA, 92673-6218

Description

Heartsine Samaritan AED Defibrillators - Models: SAM 001, SAM 002, SAM 003. Serial numbers within range of 1270 - 2324 (not all are affected).

Reason

The affected Automatic External Defibrillators (AED) annouce a fault warning and shut themselves off before a shock can be delivered to the patient. This is reportedly due to slow capacitor charging rates which are interpreted by the AED software as a fault condition.

Action

Recall was initiated by telephone 2/14/2005 with follow-up fax, email or postal mailing using one of several customized letters for distributors or customers. Customers will be sent an upgrade kit by overnite mail and forms to confirm upgrade.

Distribution

Nationwide to distributors and user facilities. Also to Argentina, Austria, Belgium, Finland, France, Germany, Hong Kong, Ireland, Israel, Italy, Japan, Kuwait, Netherlands, Portugal, Puerto Rico, Singapore, Spain, Switzerland, Thailand, United Kingdom.

Quantity

672