VITEK 2 Systems Software Version 9.01 Update Kit.

Recall: Z-0639-2019 · Initiated November 13, 2018

Recall

Recall Number
Z-0639-2019
Event Number
81589
Firm
bioMerieux, Inc.
FEI Number
3002769706
Product Code
LON
Status
Terminated
Root Cause
Software design
Initiated
November 13, 2018
Terminated
July 7, 2021
Address
100 Rodolphe St, Durham, NC, 27712-9402

Description

VITEK 2 Systems Software Version 9.01 Update Kit.

Reason

The VITEK 2 Systems Software Version 9.01 software may display incorrect organism identification information as well as the associated Antimicrobial Susceptibility Testing (AST) results when using the VITEK 2 FLEXPREP" feature.

Action

bioMerieux notified customers of the recall on about 11/13/2018 via "URGENT PRODUCT CORRECTION NOTICE." Instructions included taking the following actions: 1) Confirm the letter has been distributed to, and reviewed by, all appropriate personnel; 2) If appropriate in accordance with consignee's laboratory's Standard Operating Procedures, perform retrospective review of any patient results potentially impacted in relation to the described VITEK FLEXPREP anomaly; 3) Store this letter with the bioMrieux, Inc. VITEK 2 instrument documentation; 4) Complete and return the Acknowledgement Form in Attachment A to confirm receipt of the notice and return it to your local bioMrieux, Inc. representative. Regardless of each consignee's current use of the VITEK FLEXPREP" feature, your local Field Service Engineer has implemented the software solution on your VITEK 2 System. If you require additional assistance or have any questions, please contact your local bioMrieux, Inc. Clinical Customer Service representative at (800) 682-2666. For non-urgent inquiries, please contact the Technical Support Center by email at [email protected]. The anticipated turnaround time for email inquiries is 24-48 hours during normal business hours (7AM 7PM EST), and up to 72 hours dependent upon weekends and company observed holidays.

Distribution

Nationwide distribution to CO, IL, MO, TX.

Quantity

five (5) VITEK¿ 2 Systems Software version 9.01