Perkin Elmer WIZARD2 1-detector,3", 1000 samples. Product Code: 2480-0010 The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples
Recall
- Recall Number
- Z-0638-2016
- Event Number
- 72808
- Firm
- Perkinelmer
- FEI Number
- 3006159931
- Product Code
- JJJ
- Status
- Terminated
- Root Cause
- Nonconforming Material/Component
- Initiated
- December 8, 2015
- Posted
- January 12, 2016
- Terminated
- August 4, 2017
- Address
- 940 Winter Street, Waltham, MA, 02451
Description
Perkin Elmer WIZARD2 1-detector,3", 1000 samples. Product Code: 2480-0010 The Wizard gamma counter is intended to detect and count gamma radiation emitted by clinical samples
The content of the second #023 barcode ID label in the Barcode ID Label Binder is incorrect. The second barcode ID label #023 symbol incorrectly identifies as #024 when scanned. If measurement protocols have been programmed for both ID #023 and #024, the protocol associated with barcode ID #024 is executed. If the error is undetected, the Gamma Counter may produce erroneous results.
Perkin Elmer sent an Urgent Medical Device Correction letter dated December 7, 2015 to affected customers. The letter identified the affected products, problems and actions to be taken. Customers were instructed to destroy all affected products in their possession. Customers were asked to complete the enclosed Correction Response Form and return it by scan and e-mail to [email protected] . Corrected barcode labels ID #023 will be provided upon request and return of the completed Correction Response Form to PerkinElmer. For further information contact your local PerkinElmer representatives..
Worldwide Distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BOLIVIA, BRAZIL, CANADA, CHINA, DENMARK, FINLAND, FRANCE, GERMANY, GREAT BRITAIN, GREECE, HONG KONG, INDIA, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, KUWAIT, MYANMAR, NETHERLANDS, NEW ZEALAND, NORWAY, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, and THAILAND.
188 units