VIASYS VELA Ventilator, Catalog Number-16186; Viasys Respiratory Care, Inc.

Recall: Z-0638-2007 · Initiated January 31, 2007

Recall

Recall Number
Z-0638-2007
Event Number
37318
Firm
Viasys Respiratory Care, Inc.dba Bird Products
FEI Number
2021710
Product Code
CBK
Status
Terminated
Root Cause
Other
Initiated
January 31, 2007
Posted
March 20, 2007
Terminated
September 18, 2008
Address
1100 Bird Center Dr, Palm Springs, CA, 92262-8000

Description

VIASYS VELA Ventilator, Catalog Number-16186; Viasys Respiratory Care, Inc.

Reason

The graphical user interface may become slow in its response and in some cases non-responsive. Under some circumstances this situation may cause the ventilator to delay the annunciation of alarms.

Action

On January 31, 2007, a product notification letter was mailed to all customers in the United States via overnight carrier (Federal Express). Consignees are informed in the notification letter that service calls are in process of being scheduled to update domestic consignees. On January 31, 2007, a product notification kit (including a product notification letter, download software, download instruction sheets and installation verification fax-back sheets) were mailed to all international customers via register mail. Consignees are informed in the notification letter that replacement software, replacement instructions, and installation verification forms are included in their recall package. Domestic Consignees will have product corrected by VIASYS service personnel. Self address postage paid envelopes will be provided with notifications to aid in effectiveness checks

Distribution

Worldwide Distribution ---USA and countries of Bangladesh, Bosnia, Brazil, China, Czech Republic, Ecuador, Estonia, Finland, France, Guatemala, Hong Kong, India, Israel, Italy, Japan, Latvia, Lithuania, Malaysia, Morocco, Netherlands, Portugal, Russian Federation, Taiwan, Thailand, Turkey & Vietnam.

Quantity

878