VIASYS VELA Ventilator, Catalog Number-16186; Viasys Respiratory Care, Inc.
Recall
- Recall Number
- Z-0638-2007
- Event Number
- 37318
- Firm
- Viasys Respiratory Care, Inc.dba Bird Products
- FEI Number
- 2021710
- Product Code
- CBK
- Status
- Terminated
- Root Cause
- Other
- Initiated
- January 31, 2007
- Posted
- March 20, 2007
- Terminated
- September 18, 2008
- Address
- 1100 Bird Center Dr, Palm Springs, CA, 92262-8000
Description
VIASYS VELA Ventilator, Catalog Number-16186; Viasys Respiratory Care, Inc.
The graphical user interface may become slow in its response and in some cases non-responsive. Under some circumstances this situation may cause the ventilator to delay the annunciation of alarms.
On January 31, 2007, a product notification letter was mailed to all customers in the United States via overnight carrier (Federal Express). Consignees are informed in the notification letter that service calls are in process of being scheduled to update domestic consignees. On January 31, 2007, a product notification kit (including a product notification letter, download software, download instruction sheets and installation verification fax-back sheets) were mailed to all international customers via register mail. Consignees are informed in the notification letter that replacement software, replacement instructions, and installation verification forms are included in their recall package. Domestic Consignees will have product corrected by VIASYS service personnel. Self address postage paid envelopes will be provided with notifications to aid in effectiveness checks
Worldwide Distribution ---USA and countries of Bangladesh, Bosnia, Brazil, China, Czech Republic, Ecuador, Estonia, Finland, France, Guatemala, Hong Kong, India, Israel, Italy, Japan, Latvia, Lithuania, Malaysia, Morocco, Netherlands, Portugal, Russian Federation, Taiwan, Thailand, Turkey & Vietnam.