FDA Recall Terminated

Liaison 25 OH Vitamin D Assay, Manufacturer part #310900, Lot #114668

Recall: Z-0638-05 · Initiated February 14, 2005

Recall

Recall Number
Z-0638-05
Event Number
31228
Firm
Diasorin Inc.
FEI Number
2182595
Product Code
MRG
Status
Terminated
Root Cause
Other
Initiated
February 14, 2005
Posted
March 18, 2005
Terminated
July 22, 2006
Address
1951 Northwestern Ave S, Stillwater, MN, 55082-7536

Description

Liaison 25 OH Vitamin D Assay, Manufacturer part #310900, Lot #114668

Reason

An internal inspection indicates that Calibrator 2 and Conjugate are switched in their positions in the integral. This will result in Calibrator 1 RLU's similar to that of background RLU, and Calibrator 2 RLU's near expected range. A curve can not be calculated from this data.

Action

DiaSorin Customer Notification dated 02/14/05 was sent to customers instructing them to destroy all kits from the affected lot number. Customers were also asked to complete and return a form indicating how many kits were destroyed.

Distribution

MA, Spain, Germany & Belgium

Quantity

31 kits