LIAISON 25 OH Vitamin D Assay, Manufacturer part # 310900, Lot # 114439E.

Recall: Z-0637-05 · Initiated February 14, 2005

Recall

Recall Number
Z-0637-05
Event Number
31167
Firm
Diasorin Inc.
FEI Number
2182595
Product Code
MRG
Status
Terminated
Root Cause
Other
Initiated
February 14, 2005
Posted
March 18, 2005
Terminated
July 22, 2006
Address
1951 Northwestern Ave S, Stillwater, MN, 55082-7536

Description

LIAISON 25 OH Vitamin D Assay, Manufacturer part # 310900, Lot # 114439E.

Reason

Results of an internal inspection show Lot 114439E to have calibrator 1 and Calibrator 2 switched in their position in the integral. This results in an inability to calculate a curve, due to the reversal of calibrator values.

Action

DiaSorin Customer Notification dated 02/14/05 was sent to customers instructing them to destroy all kits from the affected lot number. Customers were also asked to complete and return a form indicating how many kits were destroyed.

Distribution

MI, NJ, Germany, Spain and Belgium

Quantity

131