FDA Recall
Terminated
Cytomics RXP Software, Version 1.0, Part Numbers: 175488 6418489 175260 175261 175262 175263 175264 175265
Recall: Z-0637-04
·
Initiated February 13, 2004
Recall
- Recall Number
- Z-0637-04
- Event Number
- 28379
- Firm
- Beckman Coulter Inc
- FEI Number
- 2050012
- Product Code
- GKL
- Status
- Terminated
- Root Cause
- Other
- Initiated
- February 13, 2004
- Posted
- July 20, 2004
- Terminated
- May 8, 2012
- Address
- 200 S Kraemer Blvd, Brea, CA, 92821-6208
Description
Cytomics RXP Software, Version 1.0, Part Numbers: 175488 6418489 175260 175261 175262 175263 175264 175265
Reason
Software anomaly. Sample ID and the Run Date may become fixed within a header on the FlowPAGE printout. All samples subsequently run, with results printed using the FlowPAGE print format, will have the fixed Sample ID and Run Date, rather than the correct Sample ID and Run Date.
Action
Firm sent modified operating instructions to users on 13 February 2004.
Distribution
Nationwide and Canada
Quantity
390