Pentax Video Colonoscope is intended to be used with a PENTAX video processor (including light source), documentation equipment, Monitor, EndoTherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract including the anus, rectum, sigmoid colon, colon and ileocecal valve.
Recall
- Recall Number
- Z-0636-2017
- Event Number
- 75306
- Firm
- Pentax of America Inc
- FEI Number
- 1000080301
- Product Code
- FDF
- Status
- Terminated
- Root Cause
- Other
- Initiated
- August 15, 2016
- Posted
- November 19, 2016
- Terminated
- March 8, 2018
- Address
- 3 Paragon Dr, Montvale, NJ, 07645-1782
Description
Pentax Video Colonoscope is intended to be used with a PENTAX video processor (including light source), documentation equipment, Monitor, EndoTherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract including the anus, rectum, sigmoid colon, colon and ileocecal valve.
Pentax is initiating the product correction of several model numbers of Pentax's Colonoscope that were shipped between the clearance date of April 2014 and September 2015 because they contain the operation and reprocessing IFU's that were in use prior to the 2014 clearances and therefore did not include the cleared operation IFU, the updated reprocessing IFU or the dedicated reprocessing adaptor (for applicable models).
Pentax sent as "Urgent Field Correction" and Field Correction Response Form dated August 15, 2016 to their affected customers.
US Nationwide and Japan
5,090 (US) and 14,209 (OUS)