FDA Recall Terminated

Pentax Video Colonoscope is intended to be used with a PENTAX video processor (including light source), documentation equipment, Monitor, EndoTherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract including the anus, rectum, sigmoid colon, colon and ileocecal valve.

Recall: Z-0636-2017 · Initiated August 15, 2016

Recall

Recall Number
Z-0636-2017
Event Number
75306
Firm
Pentax of America Inc
FEI Number
1000080301
Product Code
FDF
Status
Terminated
Root Cause
Other
Initiated
August 15, 2016
Posted
November 19, 2016
Terminated
March 8, 2018
Address
3 Paragon Dr, Montvale, NJ, 07645-1782

Description

Pentax Video Colonoscope is intended to be used with a PENTAX video processor (including light source), documentation equipment, Monitor, EndoTherapy Device such as a Biopsy Forceps, and other ancillary equipment for endoscopy and endoscopic surgery within the lower digestive tract including the anus, rectum, sigmoid colon, colon and ileocecal valve.

Reason

Pentax is initiating the product correction of several model numbers of Pentax's Colonoscope that were shipped between the clearance date of April 2014 and September 2015 because they contain the operation and reprocessing IFU's that were in use prior to the 2014 clearances and therefore did not include the cleared operation IFU, the updated reprocessing IFU or the dedicated reprocessing adaptor (for applicable models).

Action

Pentax sent as "Urgent Field Correction" and Field Correction Response Form dated August 15, 2016 to their affected customers.

Distribution

US Nationwide and Japan

Quantity

5,090 (US) and 14,209 (OUS)