FDA Recall Terminated

Gyrus ACMI PKS Seal Open Forceps, model 915000PK, 9-3/4" Curved Jaw. Sterile EO. The device is intended to be used with the G400 electrosurgical generator. Gyrus ACMI, Inc., 136 Turnpike Road, Southborough, MA 01772-2104 USA.

Recall: Z-0635-2010 · Initiated August 18, 2009

Recall

Recall Number
Z-0635-2010
Event Number
53890
Firm
Gyrus ACMI Corporation
FEI Number
3003790304
Product Code
GEI
Status
Terminated
Root Cause
Environmental control
Initiated
August 18, 2009
Posted
January 14, 2010
Terminated
March 6, 2012
Address
136 Turnpike Rd, Southborough, MA, 01772-2118

Description

Gyrus ACMI PKS Seal Open Forceps, model 915000PK, 9-3/4" Curved Jaw. Sterile EO. The device is intended to be used with the G400 electrosurgical generator. Gyrus ACMI, Inc., 136 Turnpike Road, Southborough, MA 01772-2104 USA.

Reason

Gyrus ACMI received complaints of a problem with 3103PK, 3104PK, 3105PK, 915000PK, 915005PK, 915010PK - Gyrus ACMI PKS Seal Open Forceps. The metal shim between the jaws may detach during the procedure and fall into the surgical field. The detached shim may be undetected and be left behind in the patient. This circumstance could subsequently cause complications. Further use of this product should cease, affected product should be removed from stockrooms and returned to Gyrus ACMI.

Action

An Urgent Medical Device Recall letter, dated 08/15/09 was sent to customers notifying them of the issue. The letter identified affected product, asked that they cease further use, remove affected devices from stockrooms, contact Gyrus ACMI customer service for a return goods authorization and return product to Gyrus ACMI at no charge. A reply form is also asked to be returned to Gyrus ACMI via fax. Questions and requests for a Return Goods Authorization are directed to Gyrus ACMI customer service at 1-888-524-7266.

Distribution

US and International distribution. OUS to Include: Australia, Belgium, Bolivia, Bulgaria, Canada, Chile, China, Columbia, Croatia, Czech, Denmark, Ecuador, Estonia, Finland, France, Germany, Greece , Hong Kong, India, Ireland, Israel, Italy, Japan, Korea, Malaysia, Mexico, Netherlands, Panama, Puerto Rico, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom.

Quantity

93,665 total (63,418 US, 30,247 OUS)