FDA Recall Terminated

The Propaq LT Series monitors. Model numbers 802LTAN, 802LT0N, and 802LTRN. These devices are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO2. The most likely locations for patients to be monitored by these devices are hospital general medical-surgical, telemetry, and intermediate care floors, hospital emergency departments, transport, emergency medical services, and other healthcare applications. intended to be used by clinicians and medically qualified personnel for single or multiparameter vital signs monitoring of ambulatory and nonambulatory patients, including neonate, pediatric and adult patients. These devices are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO2. The most likely locations for patients to be monitored by these devices are hospital general medicalsurgical, telemetry, and intermediate care floors, hospital emergency departments, transport, emergency medical services, and other healthcare applications. The monitors can be used as stand-alone devices or as devices networked to an Acuity Central Station through wireless communication over a Welch Allyn FlexNet network.

Recall: Z-0633-2014 · Initiated November 21, 2013

Recall

Recall Number
Z-0633-2014
Event Number
66919
Firm
Welch Allyn Protocol, Inc
FEI Number
3023750
Product Code
MWI
Status
Terminated
Root Cause
Device Design
Initiated
November 21, 2013
Posted
January 7, 2014
Terminated
February 17, 2015
Address
8500 Sw Creekside Pl, Beaverton, OR, 97008

Description

The Propaq LT Series monitors. Model numbers 802LTAN, 802LT0N, and 802LTRN. These devices are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO2. The most likely locations for patients to be monitored by these devices are hospital general medical-surgical, telemetry, and intermediate care floors, hospital emergency departments, transport, emergency medical services, and other healthcare applications. intended to be used by clinicians and medically qualified personnel for single or multiparameter vital signs monitoring of ambulatory and nonambulatory patients, including neonate, pediatric and adult patients. These devices are indicated for ECG, noninvasive blood pressure (NIBP), respiration and SpO2. The most likely locations for patients to be monitored by these devices are hospital general medicalsurgical, telemetry, and intermediate care floors, hospital emergency departments, transport, emergency medical services, and other healthcare applications. The monitors can be used as stand-alone devices or as devices networked to an Acuity Central Station through wireless communication over a Welch Allyn FlexNet network.

Reason

Welch Allyn will update Propaq 802 Series Vital Signs Monitor, models are 802LTAN, 802LTON and 802LTR with software because the fault conditions relating to hardware or software failure or electrical noise caused by proximal defibrillation activity could, in rare instances, potentially interrupt device operations.

Action

Welch Allyn began sending out the Urgent Medical Device Correction letter, dated December 03, 2013, to consignees on December 10, 2013. The letter identified the product, the problem, and the action to be taken by the customer. Customer Actions: In the unlikely event that operation of your Propaq LT is interrupted and the display becomes blank or white, pressing the power (on/off) button will restart the device and quickly bring it back to full operational capacity. This notice should be passed on to all those who need to be aware within your organization, or to any organization where the potentially affected devices have been transferred. Welch Allyn is in the process of developing a software installation service tool that will enable installation of the software update by your biomed, IT staff or other technically qualified personnel. Upon release, the service tool will be made available to you via a download from Welch Allyn's website at www.welchallyn.com/PropaqltUpgrade or other media. In the meantime and on an on-going basis, the software update will be included if you schedule routine or other service with Welch Allyn or an authorized service provider. Customers with questions were instructed to call Welch Allyn Technical Support at 1-800-535-6663. For questions regarding this recall call 503-530-7935.

Distribution

Worldwide Distribution - USA (nationwide) and Internationally to United Arab Emirates, Argentina, Australia, Belgium, Brazil, Canada, Switzerland, Chile, China, Columbia, Costa Rica, Czech Republic, Germany, Denmark, Ecuador, Spain, France, United Kingdom, Greece, Hong Kong,Taiwan, Ireland, Italy, Japan, Korea, Republic of, Kuwait, Lithuania, Luxemborg, Mexico, Malaysia, Mozambique, Namibia, Netherlands, Poland, Portugal, Qatar, Russia, Saudi Arabia, Singapore, Sierra Leone, El Salvador, Thailand, Turkey, Uganda, Vietnam, Kosovo, and South Africa.

Quantity

6003 units total (1715 units in the US and 4288 outside the US)