FDA Recall Terminated

ETI-MP IgM Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative detection of specific IgM antibodies to Mycoplasma pneumoniae in human serum. Test kit for 192 determinations. Catalog No. 4700. For In Vitro Diagnostic use. manufactured for DiaSorin Inc., 1950 Northwestern Avenue P.O. Box 285, Stillwater, MN 55082-0285

Recall: Z-0633-05 · Initiated October 20, 2004

Recall

Recall Number
Z-0633-05
Event Number
30930
Firm
Diasorin Inc.
FEI Number
2182595
Product Code
LAZ
Status
Terminated
Root Cause
Other
Initiated
October 20, 2004
Posted
March 18, 2005
Terminated
July 22, 2006
Address
1951 Northwestern Ave S, Stillwater, MN, 55082-7536

Description

ETI-MP IgM Enzyme-Linked Immunosorbent Assay (ELISA) for the qualitative detection of specific IgM antibodies to Mycoplasma pneumoniae in human serum. Test kit for 192 determinations. Catalog No. 4700. For In Vitro Diagnostic use. manufactured for DiaSorin Inc., 1950 Northwestern Avenue P.O. Box 285, Stillwater, MN 55082-0285

Reason

2 lots of ETI-MP IgM assay kits showed a decline in Optical Density (OD) reading of the positive control and calibrators, P10 and P50. The decline in OD of the positive control and calibrators (P10 & P50) would cause the assay validation criteria to fail.

Action

Customers were sent notification letters. Recommended to detroy the remaining kits and requested a form be returned which indicates the number of kits destroyed so they can arrange replacement.

Distribution

MA, OH & IL