arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin, Creatinine. For the in vitro quantitative determination of calcium, blood urea nitrogen (BUN), glucose, albumin, and creatinine.
Recall
- Recall Number
- Z-0631-2016
- Event Number
- 72860
- Firm
- Arkray Factory USA, Inc.
- FEI Number
- 1832816
- Product Code
- CGA
- Status
- Terminated
- Root Cause
- Component design/selection
- Initiated
- December 18, 2015
- Posted
- January 15, 2016
- Terminated
- January 4, 2017
- Address
- 5182 W 76th St, Minneapolis, MN, 55439-2900
Description
arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin, Creatinine. For the in vitro quantitative determination of calcium, blood urea nitrogen (BUN), glucose, albumin, and creatinine.
One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.
The firm sent an Urgent Medical Device Recall letter to customers beginning 12/18/2015. The letter identified affected product, stated the reason for recall and asked for unused product to be returned. For instructions on returning unused product contact Lesty Jaranilla 1-877-538-8872 x3243.
Distributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN.
11 boxes (25 foil packaged test strips per box)