STAT 2 I.V. Controller, Catalog/REF No. S2, ASSEMBLED IN MEXICO, CONMED CORPORATION, 525 FRENCH ROAD, UTICA, NY 13502 --- This is a disposable device with a sterile, nonpyrogenic fluid path. For controlling the infusion of Intravenous fluids into the human body.

Recall: Z-0629-2012 · Initiated April 15, 2011

Recall

Recall Number
Z-0629-2012
Event Number
60702
Firm
ConMed Corporation
FEI Number
3001451571
Product Code
LDR
Status
Terminated
Root Cause
Mixed-up of materials/components
Initiated
April 15, 2011
Posted
January 11, 2012
Terminated
January 15, 2013
Address
525 French Road, Utica, NY, 13502

Description

STAT 2 I.V. Controller, Catalog/REF No. S2, ASSEMBLED IN MEXICO, CONMED CORPORATION, 525 FRENCH ROAD, UTICA, NY 13502 --- This is a disposable device with a sterile, nonpyrogenic fluid path. For controlling the infusion of Intravenous fluids into the human body.

Reason

ConMed received complaints of some units exhibiting inaccurate or inconsistent flow rates. It was determined that certain units were assembled with an incorrect component which could result in the devices exhibiting a different flow rate and might reduce the ability of the user to regulate the device at low flow rates.

Action

ConMed Corporation sent Urgent Medical Device Recall Letters and Response Forms dated April 14, 2011, via UPS Priority Mail pouches to all domestic consignees. The letter identified the product, the problem, and the action to be taken by the consignee. The international consignees were sent Urgent Medical Device Recall Letters and Response Forms dated April 22, 2011, via FedEx International Priority. Users were instructed to immediately stop use of the device and return it to ConMed Corporation. --- RECALL EXPANSION: Consignees were also instructed to contact all of those organization within their facility and any other facilities that they may have supplied or given these affected products to. ConMed Corp. sent out Amended Urgent Medical Device Recall Letters and Response Forms dated May 10, 2011, to the 4 consignees via UPS PriorIty Mail for the expansion of the recall for Catalog No. S2-12N, Lot 1006234. For any questions call 315-624-3237 or email [email protected].

Distribution

Worldwide Distribution - USA (nationwide) and the countries of Canada, Panama and India