FDA Recall Terminated

Accu-Chek Connect Diabetes Management App, Catalog number 07562462001 / GTIN number 00365702700000 & Catalog number 07250452001 / GTIN number 00365702700017

Recall: Z-0625-2018 · Initiated June 8, 2017

Recall

Recall Number
Z-0625-2018
Event Number
79001
Firm
Roche Diabetes Care, Inc.
FEI Number
3011393376
Product Code
NDC
Status
Terminated
Root Cause
Other
Initiated
June 8, 2017
Terminated
July 27, 2018
Address
9115 Hague Rd, Indianapolis, IN, 46256-1025

Description

Accu-Chek Connect Diabetes Management App, Catalog number 07562462001 / GTIN number 00365702700000 & Catalog number 07250452001 / GTIN number 00365702700017

Reason

Certain iOS and Android App versions contain a program error (bug) in the Bolus Advisor feature. Due to a software bug, when the OS region of the phone setting is changed, the unit of measure within the app may unexpectedly change. This creates a risk the app might not transfer the blood glucose result or the user might not correctly input numerical values for carbohydrate used for bolus advice.

Action

On 06/13/2017, Roche Diabetes Care, Inc. made the Urgent Medical Device Correction notice available to their customers through the Accu-Chek.com website and through the Accu-Chek Connect microsite. Patients are instructed to: Update to version 2.1.1 of the Accu-Chek Connect app Clear all data from the Accu-Chek blood glucose meter following the steps below, before pairing with the Accu-Chek Connect app for the first time. This clearing of the records in the memory can be accomplished on the PC or Mac by using Accu-Chek Connect Device Link. A micro-USB cable and an Accu-Chek Connect Online account are required. Also ensure that the date and time are set correctly on the blood glucose meter before pairing with the app. Failure to follow these directions might impact first time use of the apps Bolus Advisor feature. Steps to clear data from Accu-Chek meter in Connect Online: 1. Create or login to your Connect Online account on your PC or Mac. 2. Download Accu-Chek Device Link 3. Right-click on Device Link icon in system tray or dock. 4. Select Settings >Prompt after every successful upload under the Delete Patient Results from Device heading. 5. Connect your meter via micro-USB cable. 6. Follow the onscreen instructions to upload your data and clear your meter. If you use the apps Bolus Advisor feature, please carefully review and manage your active insulin and insulin bolus amounts for the first 8 hours after initial installation of the app and pairing your meter. This ensures the software will establish an accurate history for use in future bolus recommendations. They are advised to contact Accu-Chek Customer Care at 1-800-688-4578 if there are additional questions.

Distribution

US Nationwide

Quantity

67,040 unique users of version 2.1.0