FDA Recall Open, Classified

SEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use with in vitro assays for determination of IgM antibodies to Cytomegalovirus (CMV), IgM antibodies to Rubella Virus, IgM antibodies to Toxoplasma gondii, IgM antibodies to Herpes Simplex Virus Type 1 and 2. Catalog Number: SR10349

Recall: Z-0624-2024 · Initiated November 6, 2023

Recall

Recall Number
Z-0624-2024
Event Number
93543
Firm
Randox Laboratories Ltd. 55 Diamond Road Crumlin (North) Ireland
FEI Number
1000361607
Product Code
OHQ
Status
Open, Classified
Root Cause
Process control
Initiated
November 6, 2023
Posted
January 2, 2024

Description

SEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use with in vitro assays for determination of IgM antibodies to Cytomegalovirus (CMV), IgM antibodies to Rubella Virus, IgM antibodies to Toxoplasma gondii, IgM antibodies to Herpes Simplex Virus Type 1 and 2. Catalog Number: SR10349

Reason

A decrease in the reactivity of HSV Type 1/2 lgM in the Serology ToRCH lgM Positive Control when tested on the DiaSorin Liaison XL.

Action

Randox issued Urgent Medical Device Correction Letter on 11/6/23 to the US Distributor Randox Laboratories-US, Ltd. , Kearneysville, West Virginia, for issuance to the user level. Letter states reason for recall, health risk and action to take: As the control material is still testing positive for HSV Type 1/2 lgM, there is no risk to health. If the control did test negative for this analyte, the affected run should be discarded, and the samples re-analysed. Action to be taken: " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to [email protected] within five working days. " Please discard all copies of the IFU and download the latest version from www.randox.com.alth risk and action to take: (

Distribution

NY, WV

Quantity

1 unit