FDA Recall Terminated

Cmax Surgical Tables, Model #'s: Cmax110, Cmax 220 manufactured by Steris Corporation in Montgomery, AL

Recall: Z-0624-2011 · Initiated August 12, 2010

Recall

Recall Number
Z-0624-2011
Event Number
56659
Firm
Steris Corporation
FEI Number
1527821
Product Code
FQO
Status
Terminated
Root Cause
Other
Initiated
August 12, 2010
Posted
December 16, 2010
Terminated
September 27, 2012
Address
5960 Heisley Rd, Mentor, OH, 44060

Description

Cmax Surgical Tables, Model #'s: Cmax110, Cmax 220 manufactured by Steris Corporation in Montgomery, AL

Reason

Cmax Surgical Tables may lose calibration if the table is operated only on battery power and if the table is in a low battery charge condition. This could result in the table moving into an unexpected tilt position when the Level button is depressed on the hand control keypad.

Action

Steris issued Urgent Field Correction Notice letters dated August 26, 2010 to each affected consignee notifying them of the voluntary field correction. A Steris Representative will contact customers to arrange for STERIS to upgrade the software in the table. For further information about this corrective action, contact Steris at 1 800 333-8828.

Distribution

Worldwide Distribution: USA in the following states: AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WV, and in the countries of Australia, Canada, China, Columbia, Dominican Republic, India, Israel, Mexico, New Zealand, Saudi Arabia, Singapore, South Korea and Taiwan.

Quantity

1,635 tables