FDA Recall Terminated

Space Pump Pole Clamps for the Infusomat Space Pump, model number 8713130. The Pole Clamp is an accessory to Space Pumps designed to support 1 to 3 pumps on a pole.

Recall: Z-0623-2013 · Initiated March 23, 2012

Recall

Recall Number
Z-0623-2013
Event Number
63862
Firm
B. Braun Medical, Inc.
FEI Number
2523676
Product Code
FRN
Status
Terminated
Root Cause
Device Design
Initiated
March 23, 2012
Posted
December 31, 2012
Terminated
February 2, 2015
Address
901 Marcon Boulevard, Allentown, PA, 18109-9512

Description

Space Pump Pole Clamps for the Infusomat Space Pump, model number 8713130. The Pole Clamp is an accessory to Space Pumps designed to support 1 to 3 pumps on a pole.

Reason

Firm has become aware of the potential for breakage of the pole clamp locking grids, or rasters, in the Space Pole Clamps.

Action

B. Braun Medical, Inc sent an Urgent Medical Device letter dated March 23, 2012 , to all affected customers. The letter identified the product the problem and the action needed by the customer to take. B. Braun has implemented an improvement to the pump locking mechanism of t he older style pole clamps and will be upgrading any older style pole clamps that may remain within your facility. Please return the completed form from attachment 1 in the self-addressed envelope provided or fax the form to (610) 849-1197 within two (2) weeks of receipt\. If you require any additional information or have any questions, comments or concerns please contact your B. Braun Medical Customer Service at 800-627-7867.

Distribution

Worldwide Distribution--US (nationwide) including the states of AL, AZ, NC and TN.

Quantity

13,078