FDA Recall Terminated

e.cam Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Siemens Medical Systems, Inc., Nuclear Medicine Group, Hoffman Estates, IL 60195-5203

Recall: Z-0623-03 · Initiated February 10, 2003

Recall

Recall Number
Z-0623-03
Event Number
25546
Firm
Siemens Medical Solutions USA, Inc.
FEI Number
1000514697
Product Code
KPS
Status
Terminated
Root Cause
Other
Initiated
February 10, 2003
Posted
March 6, 2003
Terminated
September 15, 2004
Address
2501 N. Barrington Road, Hoffman Estates, IL, 60195-5203

Description

e.cam Emission Imaging Computed Tomography Systems used for all common nuclear medicine procedures; Siemens Medical Systems, Inc., Nuclear Medicine Group, Hoffman Estates, IL 60195-5203

Reason

Unexpected release of the radial and rotate brakes in the case of multiple simultaneous fault conditions.

Action

Siemens sent Customer Advisory letters dated 2/10/03 to all e.cam customers. The accounts were informed of the problem with unexpected camera movement when multiple fault cocnditions are presnet, causing the brake system to release and the roation of the detector assemblies, and were advised to use the emergency stop at any indication of improper camera movement. The customers were advised that they would be contacted by their Siemens Customer Service Representative to schedule a firmware upgrade that will prevent the occurrence of unintended release of the radial and rotate brakes. The upgrade will be done free of charge.

Distribution

Nationwide, United Arab Emirates, Argentina, Austria, Australia, Bosnia, Barbados, Bangladesh, Belgium, Bolivia, Brazil, Camaroon, Canada, Chile, China, Colombia, Republic of Congo, Czech Republic, Denmark, Dominican Republic, El Salvador, Equador, Egypt, France, Finland, Germany, Great Britain, Georgia, Greece, Hong Kong, Honduras, Ireland, Israel, India, Italy, Iraq, Jordan, Japan, Korea, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Mexico, Morocco, Madagascar, Myanmar, Mauritius, Malaysia, Niger, Norway, New Zealand, Oman, Peru, Philippines, Pakistan, Poland, Potugal, Russia, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Turkey, Ukraine, Uzbekistan, South Africa and the Netherlands.

Quantity

1927 units