FDA Recall Open, Classified

Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Y-TYPE BLOOD SET, 3 CARSITE LADS, 146 IN; Catalog Number: 490530.

Recall: Z-0618-2026 · Initiated October 29, 2025

Recall

Recall Number
Z-0618-2026
Event Number
97816
Firm
B Braun Medical Inc
FEI Number
2521402
Product Code
FPA
Status
Open, Classified
Root Cause
Process control
Initiated
October 29, 2025
Posted
November 26, 2025
Address
824 12th Ave, Bethlehem, PA, 18018-3524

Description

Blood Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. Y-TYPE BLOOD SET, 3 CARSITE LADS, 146 IN; Catalog Number: 490530.

Reason

Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).

Action

B Braun notified consignees on about 10/29/2025 via letter. Consignees were instructed to follow "Interim Measures for Users" as outlined in the letter for secondary infusions via gravity sets and if using secondary infusions with infusion pumps. Consignees were also instructed to inform all users and post the applicable attachment(s) in care areas where secondary infusions may be administered and where IV administration sets and pump administration sets may be stored until further notice. B. Braun will provide additional notification when users may resume their normal secondary administration process without the need for the modified workflow. Consider priming all IV administration sets and pump administration sets with normal saline or other standard solutions prior to priming with expensive medications to avoid loss of expensive medications or occupational exposure to hazardous medications in the event the set is unable to be primed. If difficulty priming is encountered, obtain a replacement device and file a customer complaint with B. Braun. Additionally, consignees were requested to ensure all pertinent employees and customers are informed, post the notification where affected products are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form."

Distribution

Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.

Quantity

33,528 units