FDA Recall Open, Classified

Hookup Accessories used with MEDIVATORS ADVANTAGE PLUS and ADVANTAGE PLUS Pass Thru Automated Endoscope Reprocessors Model # Model Description 2-8-001 HU ZERO CHANNEL NO LT 2-8-010 ASSY, HOOK-UP, OLYMPUS 2-8-014 HOOKUP, OPTIM LT 2-8-030 ASSY, HOOKUP, PENTAX-0K 2-8-040 ASSY, HKUP STORZ 0K 2-8-070 ASSY, HU VISION SCIENCE 2-8-112 ASSY, HOOKUP OLMPUS 1K 2-8-120 ASSY, HOOKUP FUJI 1K 2-8-210 ASSY, HOOK-UP OLYMPUS 2K 2-8-213 ASSY, HOOKUP, OLYMPUS-2K 2-8-218 ASSY HOOKUP OLY 2K 2-8-236 ASSY, HOOKUP FUJINON 2K 2-8-244 ASSY, HOOKUP, STORZ 2K 2-8-245 ASSY, HOOK-UP, STORZ 2K 2-8-246 ASSY, HKUP, PENTAX-2K 2-8-260 ASSY, HOOK-UP WOLF 2-K 2-8-280 ASSY HOOKUP 2-8-312 ASSY, HOOK-UP OLYMPUS-3K 2-8-330 ASSY, HOOK-UP, PENTAX-3K 2-8-332 ASSEMBLY, HOOK-UP, PENTAX 2-8-334 ASSEMBLY, HOOK-UP, PENTAX 2-8-335 ASSY, HOOKUP, OLYMPUS, 3K 2-8-336 ASSY,HKUP FUJI 70 BRONC 2-8-338 ASSY, HU, FUJI BRONCH U/S 2-8-421 ASSY, HOOKUP, FUJIFILM 2-8-428 ASSY, ED-580XX 2-8-431 ASSY, HOOKUP, PENTAX 2-8-432 ASSY, HOOK-UP, PENTAX 2-8-460 HOOKUP, ASSY WOLF 4K 2-8-511 ASSY, HOOK-UP, OLY 2-8-513 ASSY HOOKUP OLY 4K ESOPHA 2-8-522 ASSY, FUJINON 4K-JET 2-8-523 ASSY, HKUP, FUJI 4K SER 5 2-8-526 ASSY, HOOKUP FUJI 4K+JET 2-8-527 ASSY, FUJI 700 NO FWJ 2-8-528 HOOK, EI-580BT 2-8-534 ASSY, HOOK-UP, PEN-4K+JET 2-8-537 ASSY, HKUP, PTX90 4K+JET 2-8-539 HOOK, PTX I20 FWJ 2-8-540 ASSY-HOOK-UP,STORZ,5K 2-8-611 ASSY, HOOK-UP, OLY 2-8-612 ASSY, HOOKUP, PENTAX 2-8-614 ASM HU OLY TGF-UC180J 2-8-617 ASSY, HKUP, OLY 2 BIOPSY 2-8-623 ASSY,HKUP,FUJI-5 4K+JET 2-8-625 ASSY, HOOKUP, EG-580UT 2-8-626 ASSY, HOOKUP, EG-580UR 2-8-627 ASSY, HKUP, FUJI 700 SER 2-8-634 ASSY, HOOKUP PENTAX DUAL 2-8-639 HOOK, EG3X-J10 2-8-640 ASSY,HOOK-UP STORZ 6K 2-8-711 ASSY, HOOKUP, OLYMPUS 5K 2-8-718 ASSY, HOOKUP, OLY 160/180 2-8-740 ASSY, HOOK-UP, STORZ 2-8-811 ASSY HKUP OLY-ECHO+LIFT 2-8-901 KIT TEE/TOE/RUS HU ADV+ 5-8-010 ASSY, HOOKUP DBL LT 5-8-030 ASSY, HOOKUP 5-8-040 ASSY HOOKUP STORZ 0K DBL 5-8-050 ASSY, HOOKUP, DILATOR 5-8-112 HOOKUP 2-SCOPE 1-K 5-8-190 HOOK DBL, IS, ION 5-8-210 ASSY, HOOKUP 5-8-245 HOOKUP KARLSTORZ-2K 2-8-001HAN HU ZERO CHANNEL NO LT 2-8-002HAN ADVANTAGE TEST BLOCK 2-8-010HAN ASSEMBLY, HOOK-UP,OLYMPUS 2-8-030HAN ASSY,HOOKUP,PENTAX-OK 2-8-112HAN ASSY, HOOK-UP OLYMPUS 1K 2-8-160HAN ASSY, ADV HOOKUP, WOLF 2-8-210HAN ASSY, HOOK-UP OLYMPUS 2K 2-8-213HAN ASSY, HOOKUP, OLYMPUS-2K 2-8-218HAN ASSY HOOKUP OLY 2K 2-8-245HAN ASSY, HOOKUP, STORZ-2K 2-8-330HAN ASSY, HOOK-UP, PENTAX-3K 2-8-334HAN ASSEMBLY, HOOK-UP, PENTAX 2-8-335HAN ASSY,HOOKUP,OLYMPUS 2-8-336HAN ASSY,HKUP FUJI 70 BRONC 2-8-338HAN ASSY, HU, FUJI BRONCH U/S 2-8-421HAN ASSY, HOOKUP, FUJINON 2-8-428CAS HOOK, CAS, ED-580XX 2-8-428HAN HOOK, HAN, ED-580XX 2-8-431CAS ASSY, CASSETTE, PENTAX 2-8-431HAN ASSY, HOOK-UP, PENTAX 2-8-511HAN ASSY, HOOK-UP, OLY 2-8-526HAN ASSY, HOOKUP FUJI 4K+JET 2-8-536HAN HOOKUP,PENTAX,70ULT, 5K 2-8-537CAS ASSY,CASS,PTX 90 4K+JET 2-8-537HAN ASSY HKUP,PTX90 4K+JET 2-8-611HAN ASSY, HOOK-UP, OLY 2-8-626CAS ASSY, CASS, FUJINON 580UR 2-8-627HAN ASSY, HAN, FUJI 700 SER 2-8-634CAS ASSY,CASS PENTAX DUAL CH 2-8-634HAN ASSY,HOOKUP PENTAX DUAL 2-8-640HAN ASSY, HOOK-UP STORZ 6K 2-8-711HAN ASSY, HOOKUP, OLYMPUS-5K 2-8-718HAN ASSY,HOOK-UP, OLY 160/180 2-8-811HAN ASSY HKUP OLY-ECHO W/LIFT 2-8-902H HU BLOCK ULTRASOUND PROBE 5-8-030HAN ASSY, HOOKUP 5-8-050HAN ASSY,HOOKUP, DILATOR 5-8-112HAN ASSY, HOOKUP 2-SCOPE 1-K 5-8-244HAN ASSY, HKUP 2K DBL

Recall: Z-0617-2024 · Initiated November 13, 2023

Recall

Recall Number
Z-0617-2024
Event Number
93566
Firm
Steris Corporation
FEI Number
1527821
Product Code
FEB
Status
Open, Classified
Root Cause
Process control
Initiated
November 13, 2023
Posted
December 22, 2023
Address
5960 Heisley Rd, Mentor, OH, 44060-1834

Description

Hookup Accessories used with MEDIVATORS ADVANTAGE PLUS and ADVANTAGE PLUS Pass Thru Automated Endoscope Reprocessors Model # Model Description 2-8-001 HU ZERO CHANNEL NO LT 2-8-010 ASSY, HOOK-UP, OLYMPUS 2-8-014 HOOKUP, OPTIM LT 2-8-030 ASSY, HOOKUP, PENTAX-0K 2-8-040 ASSY, HKUP STORZ 0K 2-8-070 ASSY, HU VISION SCIENCE 2-8-112 ASSY, HOOKUP OLMPUS 1K 2-8-120 ASSY, HOOKUP FUJI 1K 2-8-210 ASSY, HOOK-UP OLYMPUS 2K 2-8-213 ASSY, HOOKUP, OLYMPUS-2K 2-8-218 ASSY HOOKUP OLY 2K 2-8-236 ASSY, HOOKUP FUJINON 2K 2-8-244 ASSY, HOOKUP, STORZ 2K 2-8-245 ASSY, HOOK-UP, STORZ 2K 2-8-246 ASSY, HKUP, PENTAX-2K 2-8-260 ASSY, HOOK-UP WOLF 2-K 2-8-280 ASSY HOOKUP 2-8-312 ASSY, HOOK-UP OLYMPUS-3K 2-8-330 ASSY, HOOK-UP, PENTAX-3K 2-8-332 ASSEMBLY, HOOK-UP, PENTAX 2-8-334 ASSEMBLY, HOOK-UP, PENTAX 2-8-335 ASSY, HOOKUP, OLYMPUS, 3K 2-8-336 ASSY,HKUP FUJI 70 BRONC 2-8-338 ASSY, HU, FUJI BRONCH U/S 2-8-421 ASSY, HOOKUP, FUJIFILM 2-8-428 ASSY, ED-580XX 2-8-431 ASSY, HOOKUP, PENTAX 2-8-432 ASSY, HOOK-UP, PENTAX 2-8-460 HOOKUP, ASSY WOLF 4K 2-8-511 ASSY, HOOK-UP, OLY 2-8-513 ASSY HOOKUP OLY 4K ESOPHA 2-8-522 ASSY, FUJINON 4K-JET 2-8-523 ASSY, HKUP, FUJI 4K SER 5 2-8-526 ASSY, HOOKUP FUJI 4K+JET 2-8-527 ASSY, FUJI 700 NO FWJ 2-8-528 HOOK, EI-580BT 2-8-534 ASSY, HOOK-UP, PEN-4K+JET 2-8-537 ASSY, HKUP, PTX90 4K+JET 2-8-539 HOOK, PTX I20 FWJ 2-8-540 ASSY-HOOK-UP,STORZ,5K 2-8-611 ASSY, HOOK-UP, OLY 2-8-612 ASSY, HOOKUP, PENTAX 2-8-614 ASM HU OLY TGF-UC180J 2-8-617 ASSY, HKUP, OLY 2 BIOPSY 2-8-623 ASSY,HKUP,FUJI-5 4K+JET 2-8-625 ASSY, HOOKUP, EG-580UT 2-8-626 ASSY, HOOKUP, EG-580UR 2-8-627 ASSY, HKUP, FUJI 700 SER 2-8-634 ASSY, HOOKUP PENTAX DUAL 2-8-639 HOOK, EG3X-J10 2-8-640 ASSY,HOOK-UP STORZ 6K 2-8-711 ASSY, HOOKUP, OLYMPUS 5K 2-8-718 ASSY, HOOKUP, OLY 160/180 2-8-740 ASSY, HOOK-UP, STORZ 2-8-811 ASSY HKUP OLY-ECHO+LIFT 2-8-901 KIT TEE/TOE/RUS HU ADV+ 5-8-010 ASSY, HOOKUP DBL LT 5-8-030 ASSY, HOOKUP 5-8-040 ASSY HOOKUP STORZ 0K DBL 5-8-050 ASSY, HOOKUP, DILATOR 5-8-112 HOOKUP 2-SCOPE 1-K 5-8-190 HOOK DBL, IS, ION 5-8-210 ASSY, HOOKUP 5-8-245 HOOKUP KARLSTORZ-2K 2-8-001HAN HU ZERO CHANNEL NO LT 2-8-002HAN ADVANTAGE TEST BLOCK 2-8-010HAN ASSEMBLY, HOOK-UP,OLYMPUS 2-8-030HAN ASSY,HOOKUP,PENTAX-OK 2-8-112HAN ASSY, HOOK-UP OLYMPUS 1K 2-8-160HAN ASSY, ADV HOOKUP, WOLF 2-8-210HAN ASSY, HOOK-UP OLYMPUS 2K 2-8-213HAN ASSY, HOOKUP, OLYMPUS-2K 2-8-218HAN ASSY HOOKUP OLY 2K 2-8-245HAN ASSY, HOOKUP, STORZ-2K 2-8-330HAN ASSY, HOOK-UP, PENTAX-3K 2-8-334HAN ASSEMBLY, HOOK-UP, PENTAX 2-8-335HAN ASSY,HOOKUP,OLYMPUS 2-8-336HAN ASSY,HKUP FUJI 70 BRONC 2-8-338HAN ASSY, HU, FUJI BRONCH U/S 2-8-421HAN ASSY, HOOKUP, FUJINON 2-8-428CAS HOOK, CAS, ED-580XX 2-8-428HAN HOOK, HAN, ED-580XX 2-8-431CAS ASSY, CASSETTE, PENTAX 2-8-431HAN ASSY, HOOK-UP, PENTAX 2-8-511HAN ASSY, HOOK-UP, OLY 2-8-526HAN ASSY, HOOKUP FUJI 4K+JET 2-8-536HAN HOOKUP,PENTAX,70ULT, 5K 2-8-537CAS ASSY,CASS,PTX 90 4K+JET 2-8-537HAN ASSY HKUP,PTX90 4K+JET 2-8-611HAN ASSY, HOOK-UP, OLY 2-8-626CAS ASSY, CASS, FUJINON 580UR 2-8-627HAN ASSY, HAN, FUJI 700 SER 2-8-634CAS ASSY,CASS PENTAX DUAL CH 2-8-634HAN ASSY,HOOKUP PENTAX DUAL 2-8-640HAN ASSY, HOOK-UP STORZ 6K 2-8-711HAN ASSY, HOOKUP, OLYMPUS-5K 2-8-718HAN ASSY,HOOK-UP, OLY 160/180 2-8-811HAN ASSY HKUP OLY-ECHO W/LIFT 2-8-902H HU BLOCK ULTRASOUND PROBE 5-8-030HAN ASSY, HOOKUP 5-8-050HAN ASSY,HOOKUP, DILATOR 5-8-112HAN ASSY, HOOKUP 2-SCOPE 1-K 5-8-244HAN ASSY, HKUP 2K DBL

Reason

Identification label on affected hookups could become illegible over time due to the label's ink coming off. This may potentially prevent users from being able to effectively identify the hookup and may ultimately result in postponement of patient procedures.

Action

On November 13, 2023, the firm notified customers via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to inspect their inventory for affected product. You should not attempt to use a hookup that cannot be effectively identified. Steris will provide either a replacement hookup or replacement label for all affected hookups. If you have questions regarding this matter, please contact [email protected].

Distribution

Domestic distribution nationwide. Foreign distribution to Australia Canada China Hong Kong Italy Malaysia Netherlands New Zealand Singapore Thailand

Quantity

2545 US; 992 OUS