FDA Recall Completed

Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000

Recall: Z-0616-2024 · Initiated November 15, 2023

Recall

Recall Number
Z-0616-2024
Event Number
93524
FEI Number
3008426232
Product Code
IYE
Status
Completed
Root Cause
Other
Initiated
November 15, 2023
Posted
December 22, 2023
Address
Landauer 2 Science Rd, Glenwood, IL, 60425-1531

Description

Landauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000

Reason

MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly.

Action

Landauer Issued an URGENT - Medical Device Recall notice to its consignees on 11/15/2023 via email. The notice explained the issue, potential risk, and requested return of the device.

Distribution

US and Australia, Brazil, Canada, China, France, Germany, India, Japan, Mexico, Rep of Korea, Taiwan, Thailand, Viet Nam

Quantity

584 units